BC THROMBIN REAGENT 10446636

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-11-01 for BC THROMBIN REAGENT 10446636 manufactured by Siemens Healthcare Diagnostics Products Gmbh.

Event Text Entries

[91921629] Siemens healthcare diagnostics investigated the event to determine the cause of the elevated thrombin time (tt) results on the bcs xp systems and determined that the customer was comparing two different methods for tt determination. The customer used the bc thrombin reagent on the bcs xp systems, while they used the test thrombin reagent on the sysmex ca-660 system to obtain tt results on patient samples. Siemens healthcare diagnostics has confirmed that the bc thrombin reagent kit lot 46751 (contains thrombin reagent lot 517468) and kit lot 47184 (contains thrombin reagent lot 517469) produce unexpected prolonged thrombin time (tt) results for expected normal samples, and may recover above the upper limit of normal (< 21 seconds) as stated within the ifu (instruction for use). The investigation by siemens confirmed that the reference range is shifted to an approximately 20% prolonged tt. If the reference range is not adjusted for the current affected lot, this may lead to a higher number of samples requiring follow-up in the case of slightly prolonged tt. There is a potential for misinterpreting unexpected prolonged results as unfractionated (uf) heparin contamination and some other thrombin inhibitors. An urgent field safety notice (ufsn) ph-17-018. A. Ous was sent to outside the us (ous) customers and an urgent medical device correction (umdc) ph17-018. A. Us was sent to us customers in october 2017. The ufsn and umdc remind customers that reference intervals vary from laboratory to laboratory depending on the population, the technique and reagent lot. Therefore, each laboratory must establish its own reference intervals or verify them whenever one or more of the aforementioned variables are changed. The ufsn and udmc also informs customers using the bc thrombin reagent as a screening assay for thrombin inhibition that the affected lots are more sensitive with approximately 20% prolonged/elevated tt results. This will lead to an increased rate of results above the reference range if not adjusted. Thus, the ufsn and udmc advise customers to adjust their specific reference ranges as per the application sheet for thrombin time with bc thrombin when changing to a new lot. Mdr 9610806-2017-00120, mdr 9610806-2017-00124, mdr 9610806-2017-00126, and mdr 9610806-2017-00127 were filed for the same event.
Patient Sequence No: 1, Text Type: N, H10


[91921630] Elevated thrombin time (tt) results were obtained on multiple patient samples on two bcs xp systems using bc thrombin reagent lot 517468 (kit lot 46751). These results were not reported to the physician(s). The same samples were repeated on an alternate sysmex ca-660 system using an alternate test thrombin reagent, resulting lower. The results obtained on the sysmex ca-660 system were reported to the physician. While the internal quality controls (qc) were within range for bcs xp system serial number (b)(4) and the sysmex ca-660 system, it is unknown if the internal qc were within range for bcs xp system serial number (b)(4) at the time of the event. There are no known reports of patient intervention or adverse health consequences due to the elevated tt results.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9610806-2017-00125
MDR Report Key6995315
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-11-01
Date of Report2017-11-01
Date of Event2017-07-25
Date Mfgr Received2017-10-05
Date Added to Maude2017-11-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCHRISTINA LAM
Manufacturer Street511 BENEDICT AVE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145243504
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
Manufacturer StreetEMIL-VON-BEHRING- STR. 76
Manufacturer CityMARBURG, D-35041
Manufacturer CountryGM
Manufacturer Postal CodeD-35041
Single Use3
Remedial ActionRC
Previous Use Code3
Removal Correction Number9610806-10/18/2017-003-C
Event Type3
Type of Report3

Device Details

Brand NameBC THROMBIN REAGENT
Generic NameBC THROMBIN REAGENT
Product CodeGJA
Date Received2017-11-01
Model NumberBC THROMBIN REAGENT
Catalog Number10446636
Lot Number46751
Device Expiration Date2018-09-08
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
Manufacturer AddressEMIL-VON-BEHRING-STR. 76 MARBURG D-35041 GM D-35041


Patients

Patient NumberTreatmentOutcomeDate
10 2017-11-01

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.