ABL700 ABL735 *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01 report with the FDA on 2006-04-10 for ABL700 ABL735 * manufactured by Radiometer Medical Aps.

Event Text Entries

[456271] The customer has tested the abl735 with a reference instrument (modular) and observed that abl735 measures lower bilirubin concentrations than modular. Neonates with a too high concentration of bilirubin may need light therapy or exchange transfusion. If the ctbil-values shown by the device are ower than the actual values, the measurements may affect this diagnosis resulting in delayed necessary treatment. Healthcare personnel will get other indications of the problem, because the skin is turning yellow (jaundice) but it may not possible to evaluate the degree of the problem exactly based on the colour. Measurements has been provided all indicating lower values for abl700.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002807968-2006-00023
MDR Report Key699823
Report Source01
Date Received2006-04-10
Date of Report2006-04-10
Date of Event2006-02-27
Date Mfgr Received2006-02-27
Device Manufacturer Date2003-01-01
Date Added to Maude2006-04-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer StreetAKANDEVEJ 21
Manufacturer CityBRONSHOJ DK-2700
Manufacturer CountryDA
Manufacturer PostalDK-2700
Manufacturer Phone8273057
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameABL700
Generic NameBLOOD GAS ANALYZER
Product CodeCCE
Date Received2006-04-10
Model NumberABL735
Catalog Number*
Lot Number*
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key688913
ManufacturerRADIOMETER MEDICAL APS
Manufacturer Address* BRONSHOJ DA
Baseline Brand NameABL735
Baseline Generic NameBLOOD GAS ANALYSER WITH ADDITIONAL PARAMETERS
Baseline Model NoABL735
Baseline Catalog NoNA
Baseline IDNA


Patients

Patient NumberTreatmentOutcomeDate
10 2006-04-10

© 2024 FDA.report