OMNICLOSE TROCAR SITE CLOSURE NEEDLE FCL201100

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-11-03 for OMNICLOSE TROCAR SITE CLOSURE NEEDLE FCL201100 manufactured by Unimax Medical Systems Inc.

Event Text Entries

[92211993] The dhr for the product in question was reviewed and found without any irregularities. It is unclear under what circumstances could the device cut the suture considering the structure and operation principle of the device. The structure of the grasping jaws are rounded without sharp edges. The operation principle is to grasp the suture and hold it in the loop created by the jaws instead of firmly clipping it. A review of the record indicates that there is no similar issue since the product is launched. The event will be closed. If the additional information could be obtained for further investigation, the supplemental report will be filed.
Patient Sequence No: 1, Text Type: N, H10


[92211994] When the graspers were closed and pulling the suture in, it cut through the suture instead of grasping it.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007791595-2017-00012
MDR Report Key6998588
Date Received2017-11-03
Date of Report2017-11-03
Date of Event2017-09-13
Date Added to Maude2017-11-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameOMNICLOSE TROCAR SITE CLOSURE NEEDLE
Generic NameTROCAR SITE CLOSURE
Product CodeGDF
Date Received2017-11-03
Model NumberFCL201100
Catalog NumberFCL201100
Lot Number8101705035
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerUNIMAX MEDICAL SYSTEMS INC
Manufacturer Address8F-2, NO 127, LANE 235 PAO CHIAO R, HSIN TIEN DIST. TAIPEI CITY, 23145 TW 23145


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-11-03

© 2026 FDA.report