MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2017-11-06 for COBAS E 411 IMMUNOASSAY ANALYZER E411 RACK 04775201001 manufactured by Roche Diagnostics.
[92090568]
(b)(4)
Patient Sequence No: 1, Text Type: N, H10
[92090569]
The customer received a discrepant elecsys probnp ii immunoassay result for a patient sample on the cobas e 411 immunoassay analyzer. The initial result was 9 pg/ml with a data flag and was reported outside of the laboratory. The result was questioned by the physician and the sample was repeated with a result of 2373 pg/ml. There was no adverse event. The reagent lot number was 22774802iu with and expiration date of 31-jul-2018. The field service engineer was not able to determine a root cause. The analyzer performance check was within specification.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1823260-2017-02535 |
MDR Report Key | 7002551 |
Report Source | HEALTH PROFESSIONAL,USER FACI |
Date Received | 2017-11-06 |
Date of Report | 2017-12-27 |
Date of Event | 2017-10-18 |
Date Mfgr Received | 2017-10-18 |
Date Added to Maude | 2017-11-06 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 0 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 0 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | NA MICHAEL LESLIE |
Manufacturer Street | 9115 HAGUE ROAD NA |
Manufacturer City | INDIANAPOLIS IN 46250 |
Manufacturer Country | US |
Manufacturer Postal | 46250 |
Manufacturer Phone | 3175214343 |
Manufacturer G1 | HITACHI HIGH TECH CORP. |
Manufacturer Street | 882 ICHIGE HITACHINAKA NA |
Manufacturer City | IBARAKI 312-8504 |
Manufacturer Country | JA |
Manufacturer Postal Code | 312-8504 |
Single Use | 3 |
Previous Use Code | 3 |
Removal Correction Number | NA |
Event Type | 3 |
Type of Report | 3 |
Brand Name | COBAS E 411 IMMUNOASSAY ANALYZER |
Generic Name | IMMUNOCHEMISTRY ANALYZER |
Product Code | NBC |
Date Received | 2017-11-06 |
Model Number | E411 RACK |
Catalog Number | 04775201001 |
Lot Number | NA |
ID Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 0 |
Device Event Key | 0 |
Manufacturer | ROCHE DIAGNOSTICS |
Manufacturer Address | 9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2017-11-06 |