COMP RVS TRAY +5MM CO 44MM N/A 115375

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-11-07 for COMP RVS TRAY +5MM CO 44MM N/A 115375 manufactured by Zimmer Biomet, Inc..

Event Text Entries

[92478645] Zimmer biomet complaint: (b)(4). Complaint sample was evaluated and the reported event was not confirmed. Product was found to be manufactured to design specifications as noted in dimensional analysis. (b)(4) was initiated based on similar complaints of assembly concerns dhr was reviewed and no discrepancies related to the reported event were found. Review of the complaint history determined that no further action is required as no trends were identified. Investigation results concluded that the reported event was due to positioning/alignment during the surgical procedure. There are warnings in the package insert that this type of event can occur and risks are addressed in risk documentation. A summary of the investigation was sent to the complainant conveying proper surgical technique and use of the device. If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly. Zimmer biomet will continue to monitor for trends.
Patient Sequence No: 1, Text Type: N, H10


[92478646] It was reported that the e1 poly bearing would not snap into the base.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0001822565-2017-07608
MDR Report Key7005632
Date Received2017-11-07
Date of Report2017-10-31
Date of Event2014-05-27
Date Mfgr Received2014-06-03
Device Manufacturer Date2013-05-17
Date Added to Maude2017-11-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. CHRISTINA ARNT
Manufacturer Street56 E. BELL DR.
Manufacturer CityWARSAW IN 46582
Manufacturer CountryUS
Manufacturer Postal46582
Manufacturer Phone5745273773
Manufacturer G1ZIMMER BIOMET, INC.
Manufacturer Street56 E. BELL DRIVE
Manufacturer CityWARSAW IN 46582
Manufacturer CountryUS
Manufacturer Postal Code46582
Single Use3
Previous Use Code3
Removal Correction NumberN/A
Event Type3
Type of Report0

Device Details

Brand NameCOMP RVS TRAY +5MM CO 44MM
Generic NameEXTREMITIES - IMPLANTS
Product CodeLRP
Date Received2017-11-07
Returned To Mfg2014-06-09
Model NumberN/A
Catalog Number115375
Lot Number394180
ID NumberN/A
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerZIMMER BIOMET, INC.
Manufacturer Address56 E. BELL DRIVE WARSAW IN 46582 US 46582


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2017-11-07

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.