MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-11-07 for ADVIA CENTAUR XPT manufactured by Siemens Healthcare Diagnostics Inc..
        [92324012]
A siemens customer service engineer (cse) was dispatched to the customer site. After analyzing the instrument, the cse replaced the aspirate probes and aspirate probe pinch valves. During the cleaning procedure, the cse noted that the aspirate probe 3 cuvette was overflowing. A siemens headquarters support center (hsc) specialist reviewed the information provided and stated that the background and precision data were acceptable. During the follow-up visit, the cse replaced the acid pump, degasser, and reagent 1 diluter. The cuvette overflow occurred during afp sample processing. There was a residual film around the wash 1 dispense port. This could result in wicking from droplets accumulating on the port. During a revisit to the customer site, the cse ran dry test, wet water test and wash 1 test and the relative light units were below 150. The cse determined that there were solution stain marks underneath the wash manifold near acid probe. Additionally, during the daily cleaning procedure, the aspirate probe 3 was overflowing. However, during the sample processing, no overflow was observed. The cause of the discordant, falsely elevated afp result on one patient sample is unknown. Siemens is investigating the issue. Mdr 2432235-2017-00531 was filed for the same issue for the event occurred on (b)(6) 2017.
 Patient Sequence No: 1, Text Type: N, H10
        [92324013]
A discordant, falsely elevated alpha-fetoprotein (afp) result was obtained on one patient sample on an advia centaur xpt instrument. It is unknown if the initial result was reported to the physician(s). The sample was repeated twice on the same instrument, resulting lower. The repeat results were reported to the physician(s). There are no reports of patient intervention or adverse health consequence due to the discordant, falsely elevated afp result.
 Patient Sequence No: 1, Text Type: D, B5
| Report Number | 2432235-2017-00599 | 
| MDR Report Key | 7007751 | 
| Report Source | FOREIGN,HEALTH PROFESSIONAL,U | 
| Date Received | 2017-11-07 | 
| Date of Report | 2018-07-23 | 
| Date of Event | 2017-10-09 | 
| Date Mfgr Received | 2018-05-07 | 
| Device Manufacturer Date | 2016-11-07 | 
| Date Added to Maude | 2017-11-07 | 
| Event Key | 0 | 
| Report Source Code | Manufacturer report | 
| Manufacturer Link | Y | 
| Number of Patients in Event | 0 | 
| Adverse Event Flag | 0 | 
| Product Problem Flag | 3 | 
| Reprocessed and Reused Flag | 3 | 
| Health Professional | 3 | 
| Initial Report to FDA | 3 | 
| Report to FDA | 3 | 
| Event Location | 3 | 
| Manufacturer Contact | SHWETA GULATI | 
| Manufacturer Street | 511 BENEDICT AVENUE | 
| Manufacturer City | TARRYTOWN NY 10591 | 
| Manufacturer Country | US | 
| Manufacturer Postal | 10591 | 
| Manufacturer Phone | 9145242870 | 
| Manufacturer G1 | SIEMENS HEALTHCARE DIAGNOSTICS | 
| Manufacturer Street | MANUFACTURING LIMITED REGISTRATION NUMBER: 8020888 | 
| Manufacturer City | CHAPEL LANE, SWORDS, CO., DUBLIN | 
| Manufacturer Country | EI | 
| Single Use | 3 | 
| Previous Use Code | 3 | 
| Event Type | 3 | 
| Type of Report | 3 | 
| Brand Name | ADVIA CENTAUR XPT | 
| Generic Name | IMMUNOASSAY ANALYZER | 
| Product Code | LOK | 
| Date Received | 2017-11-07 | 
| Model Number | ADVIA CENTAUR XPT | 
| Operator | HEALTH PROFESSIONAL | 
| Device Availability | N | 
| Device Age | DA | 
| Device Eval'ed by Mfgr | N | 
| Device Sequence No | 0 | 
| Device Event Key | 0 | 
| Manufacturer | SIEMENS HEALTHCARE DIAGNOSTICS INC. | 
| Manufacturer Address | 511 BENEDICT AVE TARRYTOWN NY 10591 US 10591 | 
| Brand Name | ADVIA CENTAUR XPT | 
| Generic Name | IMMUNOASSAY ANALYZER | 
| Product Code | JJE | 
| Date Received | 2017-11-07 | 
| Model Number | ADVIA CENTAUR XPT | 
| Operator | HEALTH PROFESSIONAL | 
| Device Availability | N | 
| Device Age | DA | 
| Device Eval'ed by Mfgr | N | 
| Device Sequence No | 1 | 
| Device Event Key | 0 | 
| Manufacturer | SIEMENS HEALTHCARE DIAGNOSTICS INC. | 
| Manufacturer Address | 511 BENEDICT AVE TARRYTOWN NY 10591 US 10591 | 
| Patient Number | Treatment | Outcome | Date | 
|---|---|---|---|
| 1 | 0 | 2017-11-07 |