TEC 6 PLUS

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-11-08 for TEC 6 PLUS manufactured by Datex-ohmeda, Inc..

Event Text Entries

[91447085] A gehc service representative performed a checkout of the equipment and confirmed the unit's output was low to manufacturer's specifications. The unit was returned to the manufacturing site for investigation. The unit was tested and the output was found to be high to manufacturing specifications. The reported fault was due to a damaged rotary valve. Patient information could not be obtained due to country privacy laws. The initial reporter is located outside the u. S. , and therefore this information is not provided due to country privacy laws.
Patient Sequence No: 1, Text Type: N, H10


[91447086] The hospital reported the patient woke up during a case. There was no reported patient injury.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2112667-2017-02088
MDR Report Key7009677
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-11-08
Date of Report2017-11-08
Date of Event2017-06-14
Date Mfgr Received2017-06-15
Device Manufacturer Date2012-01-03
Date Added to Maude2017-11-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactJOHN SZALINSKI
Manufacturer Street540 W. NORTHWEST HIGHWAY
Manufacturer CityBARRINGTON IL
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTEC 6 PLUS
Generic NameVAPORIZER
Product CodeCAD
Date Received2017-11-08
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerDATEX-OHMEDA, INC.
Manufacturer Address3030 OHMEDA DR, MADISON, WI US


Patients

Patient NumberTreatmentOutcomeDate
10 2017-11-08