ANCHOR BOLT - LSB STYLE UNKNOWN

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor report with the FDA on 2017-11-10 for ANCHOR BOLT - LSB STYLE UNKNOWN manufactured by Ad-tech Medical Instrument Corporation.

Event Text Entries

[92659775] As stated in the event description, (b)(6) from (b)(6) contacted ad-tech by email on (b)(6) 2017 regarding anchor bolts that broke. Two fractured anchor bolts were partially retrieved on an explant. This was a patient with non-convulsive seizures, and no overt trauma in the emu or theatre. One bolt was subsequently retrieved at craniotomy, another, in an uninvolved region, remains in situ. The products were not sent back to ad-tech for evaluation. Instead, a picture was sent showing the broken bolts. Based on the picture, the alleged deficiency in the complaint was confirmed. The investigation in still on-going.
Patient Sequence No: 1, Text Type: N, H10


[92659776] On (b)(6) 2017 a customer contacted ad-tech via email regarding anchor bolts that broke. It was stated that two fractured anchor bolts were partially retrieved on an explant. This was a patient with non-convulsive seizures, and no overt trauma in the emu or theatre. The customer also stated that one bolt was subsequently retrieved at craniotomy. Another one, in an uninvolved region, remained in situ. There is no known impact to the patient. Additional information has been requested.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2183456-2017-00010
MDR Report Key7022990
Report SourceDISTRIBUTOR
Date Received2017-11-10
Date of Report2017-11-10
Date Mfgr Received2017-10-12
Date Added to Maude2017-11-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. KATHLEEN BARLOW
Manufacturer Street1901 WILLIAM STREET
Manufacturer CityRACUNE WI 53404
Manufacturer CountryUS
Manufacturer Postal53404
Manufacturer Phone2626341555
Manufacturer G1AD-TECH MEDIAL INSTRUMENT CORPORATION
Manufacturer Street1901 WILLIAM STREET
Manufacturer CityRACINE WI 53404
Manufacturer CountryUS
Manufacturer Postal Code53404
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameANCHOR BOLT - LSB STYLE
Generic NameANCHOR BOLT - LSB STYLE
Product CodeGCZ
Date Received2017-11-10
Catalog NumberUNKNOWN
Lot NumberUNKNOWN
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerAD-TECH MEDICAL INSTRUMENT CORPORATION
Manufacturer Address1901 WILLIAM STREET RACINE WI 53404 US 53404


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-11-10

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.