CALLISTO EYE 301640-3020-350

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-11-10 for CALLISTO EYE 301640-3020-350 manufactured by Carl Zeiss Meditec Ag (oberkochen).

Event Text Entries

[91801534] Before starting a surgery with callisto eye, the position of the health care professional (hcp) must be set in relation to the patient's head (right, left, or top). The selected position is indicated on the live video screen throughout the surgery. In this particular case, the "right" position was selected in callisto eye, but the hcp was physically positioned near the top of the patient's head. This caused the 90 degree discrepancy in the displayed target axis.
Patient Sequence No: 1, Text Type: N, H10


[91801535] The health care professional (hcp) reported that a toric iol was implanted in the patient's right eye 90 degrees off the desired axis. The markerless function of the callisto eye had been used to position the iol. The patient's astigmatic outcome was about 6. 0 diopters, which resulted in poor visual acuity. The hcp decided to perform a follow-up surgery using the callisto eye to re-position the iol. After the corrective treatment, the patient's astigmatism is about 0. 25 diopters.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9615010-2017-00012
MDR Report Key7023374
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-11-10
Date of Report2017-10-13
Date of Event2017-10-10
Date Mfgr Received2017-10-13
Device Manufacturer Date2014-08-14
Date Added to Maude2017-11-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. WILLIAM GUSTAFSON
Manufacturer Street5160 HACIENDA DRIVE
Manufacturer CityDUBLIN CA 94568
Manufacturer CountryUS
Manufacturer Postal94568
Manufacturer Phone9255574689
Manufacturer G1CARL ZEISS MEDITEC AG (OBERKOCHEN)
Manufacturer StreetRUDOLF-EBER-STRASSE 11
Manufacturer CityOBERKOCHEN, BADEN-WUERTTEMBERG 73447
Manufacturer CountryGM
Manufacturer Postal Code73447
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCALLISTO EYE
Generic NameCALLISTO EYE
Product CodeNFJ
Date Received2017-11-10
Model NumberNA
Catalog Number301640-3020-350
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCARL ZEISS MEDITEC AG (OBERKOCHEN)
Manufacturer AddressRUDOLF-EBER-STRASSE 11 OBERKOCHEN, BADEN-WUERTTEMBERG 73447 GM 73447


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-11-10

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.