OLYMPUS WM-NP2 K10021610

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2017-11-13 for OLYMPUS WM-NP2 K10021610 manufactured by Keymed (medical And Industrial Equipment) Ltd..

Event Text Entries

[91898402] Initial investigation by olympus (b)(4) has confirmed that the castor on the rear-right side was disengaged. The bolt was assembled to the trolley body, and it was confirmed that the castor and the bolt were disconnected. Images provided by olympus (b)(4) show that the joint was broken. Keymed (medical and industrial equipment) ltd have requested the return of the subject device.
Patient Sequence No: 1, Text Type: N, H10


[91898403] Wm-np2 workstation is intended for use in medical facilities under the direction of a trained physician, and has been designed to be used with a range of olympus equipment to facilitate gi endoscopy, endoscopic ultrasound, respiratory and surgical endoscopic procedures. Keymed (medical and industrial equipment) ltd has been made aware of an event in (b)(6) where the workstation was moved between rooms during which the rear castor broke causing the workstation to fall against a nurse. Other staff members held the workstation to support its weight.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9611174-2017-00007
MDR Report Key7025707
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2017-11-13
Date of Report2017-12-05
Date of Event2017-10-17
Date Mfgr Received2017-12-05
Device Manufacturer Date2017-01-23
Date Added to Maude2017-11-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactMRS AIMILIA GKARAFLI
Manufacturer StreetKEYMED HOUSE, STOCK ROAD
Manufacturer CitySOUTHEND ON SEA, ESSEX
Manufacturer CountryUS
Manufacturer Phone0441702616
Manufacturer G1KEYMED (MEDICAL AND INDUSTRIAL EQUIPMENT) LTD
Manufacturer StreetKEYMED HOUSE, STOCK ROAD
Manufacturer CitySOUTHEND ON SEA, ESSEX
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPUS
Generic NameWM-NP2 ENDOSCOPY WORKSTATION
Product CodeFEM
Date Received2017-11-13
Model NumberWM-NP2
Catalog NumberK10021610
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerKEYMED (MEDICAL AND INDUSTRIAL EQUIPMENT) LTD.
Manufacturer AddressKEYMED HOUSE, STOCK ROAD SOUTHEND-ON-SEA, ESSEX US


Patients

Patient NumberTreatmentOutcomeDate
10 2017-11-13