IMRIS T2X OR TABLE V2 (ORT100) 110470-000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-11-13 for IMRIS T2X OR TABLE V2 (ORT100) 110470-000 manufactured by Imris - Deerfield Imaging, Inc..

Event Text Entries

[92906679] The drift issue was isolated by the service field engineer to the roll cylinder and the flow control valves. They were replaced, the hydraulic system was flushed and bled of any air, and two 8-hour drift tests were conducted to verify proper function of the table.
Patient Sequence No: 1, Text Type: N, H10


[92906680] The customer stated that the ort100 table exhibited roll drift issues. It was later discovered that a patient had been involved in the incident when the drift in the roll cylinder was observed. The table was re-leveled by or staff and the procedure was completed successfully. There was no patient injury.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3010326005-2017-00009
MDR Report Key7026574
Date Received2017-11-13
Date of Report2017-11-13
Date of Event2017-10-02
Date Mfgr Received2017-10-02
Device Manufacturer Date2010-09-27
Date Added to Maude2017-11-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. SANJAY SHAH
Manufacturer Street5101 SHADY OAK ROAD
Manufacturer CityMINNETONKA MN 553434100
Manufacturer CountryUS
Manufacturer Postal553434100
Manufacturer Phone7632036380
Manufacturer G1IMRIS - DEERFIELD IMAGING, INC.
Manufacturer Street5101 SHADY OAK ROAD
Manufacturer CityMINNETONKA MN 553434100
Manufacturer CountryUS
Manufacturer Postal Code553434100
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameIMRIS T2X OR TABLE V2 (ORT100)
Generic NameOPERATING ROOM TABLE, PRODUCT CODE KXJ
Product CodeKXJ
Date Received2017-11-13
Model NumberORT100
Catalog Number110470-000
OperatorNURSE
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerIMRIS - DEERFIELD IMAGING, INC.
Manufacturer Address5101 SHADY OAK ROAD MINNETONKA MN 553434100 US 553434100


Patients

Patient NumberTreatmentOutcomeDate
10 2017-11-13

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.