MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer,foreign report with the FDA on 2017-11-14 for PHILIPS SONICARE HX8240 manufactured by Philips Oral Healthcare.
[92017787]
Lacerations on the tongue made lead to permanent damage to a body structure.
Patient Sequence No: 1, Text Type: N, H10
[92017788]
Consumer complained the stream of the airfloss caused bleeding and a blood blister on their tongue. Medical attention sought.
Patient Sequence No: 1, Text Type: D, B5
[108298255]
Internal inspection shows no issue with unit. Charger is operating within specification. There is a warning in the dfu under important, caution on pg. 8 that states, "to avoid physical injury, do not direct the spray under the tongue or into the ear, nose, eye or other sensitive area. " the root cause of the customer's complaint could not be determined.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 3026630-2017-00280 |
MDR Report Key | 7030401 |
Report Source | CONSUMER,FOREIGN |
Date Received | 2017-11-14 |
Date of Report | 2017-11-14 |
Date of Event | 2017-10-25 |
Date Mfgr Received | 2018-01-03 |
Device Manufacturer Date | 2017-07-12 |
Date Added to Maude | 2017-11-14 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | PATIENT |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MS. NICOLE FLAVELL |
Manufacturer Street | 22100 BOTHELL EVERETT HWY |
Manufacturer City | BOTHELL WA 980413003 |
Manufacturer Country | US |
Manufacturer Postal | 980413003 |
Manufacturer Phone | 4259081225 |
Manufacturer G1 | PHILIPS ORAL HEALTHCARE |
Manufacturer Street | 22100 BOTHELL EVERETT HWY |
Manufacturer City | BOTHELL WA 980413003 |
Manufacturer Country | US |
Manufacturer Postal Code | 980413003 |
Single Use | 3 |
Previous Use Code | 3 |
Removal Correction Number | N/A |
Event Type | 3 |
Type of Report | 3 |
Brand Name | PHILIPS SONICARE |
Generic Name | AIRFLOSS 1.5 |
Product Code | EFS |
Date Received | 2017-11-14 |
Returned To Mfg | 2018-01-03 |
Model Number | HX8240 |
Operator | LAY USER/PATIENT |
Device Availability | R |
Device Age | DA |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | PHILIPS ORAL HEALTHCARE |
Manufacturer Address | 22100 BOTHELL EVERETT HWY BOTHELL WA 980413003 US 980413003 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Other | 2017-11-14 |