USA ELITE SYSTEM ROTATING CF OUTER SHEATH, 25 FR, GOLD EROS-CF25

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2017-11-14 for USA ELITE SYSTEM ROTATING CF OUTER SHEATH, 25 FR, GOLD EROS-CF25 manufactured by Gyrus Acmi, Inc.

Event Text Entries

[93194895] The device was not returned to olympus for evaluation. Olympus followed up with the user facility via telephone and in writing to obtain additional information regarding the reported event but with no result. The exact cause of the reported event cannot be determined at this time. However, the instruction manual warns users "do not use an instrument that fails to meet the criteria stated in the labeling or that has been damaged. Damage may result in the loss of the entire ceramic tip or fragments of the ceramic tip. If there is evidence of charring, burn spots, chips or cracks in the ceramic tip or surrounding area, do not use. "
Patient Sequence No: 1, Text Type: N, H10


[93194896] Olympus was informed that during an unspecified procedure, the plastic tip of the outer sheath broke off into three pieces and fell inside the patient. It was reported that the surgeon retrieved all three pieces of the plastic tip from the patient. The adjacent areas were inspected for trauma and none was noted. It is unknown if the intended procedure was completed.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2951238-2017-00723
MDR Report Key7031775
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2017-11-14
Date of Report2017-11-14
Date of Event2017-09-18
Date Mfgr Received2017-10-27
Date Added to Maude2017-11-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. CONNIE TUBERA
Manufacturer Street2400 RINGWOOD AVENUE
Manufacturer CitySAN JOSE CA 95131
Manufacturer CountryUS
Manufacturer Postal95131
Manufacturer Phone408935-512
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameUSA ELITE SYSTEM ROTATING CF OUTER SHEATH, 25 FR, GOLD
Generic NameUSA ELITE SYSTEM ROTATING CF OUTER SHEATH
Product CodeFBO
Date Received2017-11-14
Model NumberEROS-CF25
Catalog NumberEROS-CF25
Lot NumberUNKNOWN
ID NumberUDI
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerGYRUS ACMI, INC
Manufacturer Address136 TURNPIKE ROAD SOUTHBOROUGH MA 01772 US 01772


Patients

Patient NumberTreatmentOutcomeDate
10 2017-11-14

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.