SPECULUM MCO981D DEVEZE

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-11-15 for SPECULUM MCO981D DEVEZE manufactured by Integra Microfrance S.a.s..

Event Text Entries

[93177087] Integra has completed their internal investigation on november 13, 2017. Results: evaluation of returned device; the laser welding between speculum and frame is cracked. There is not fragment. A thorough observation of welding did not highlighted welding defect. Dhr review; (always include dhr review, if lot number or device not provided, explain why). No anomalies that could be associated with the complaint were observed. Complaints history; no adverse trend - first occurrence of this risk for this device. Conclusion: the origin of the crack cannot be determined with certainty. The most probable cause is an excessive constraint on the speculum.
Patient Sequence No: 1, Text Type: N, H10


[93177088] It was reported that the device seal broke. Product was in contact with the patient; however, no patient injury reported and the event lead to 45 minutes surgical delay.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2523190-2017-00136
MDR Report Key7032551
Date Received2017-11-15
Date of Report2017-11-07
Date Mfgr Received2017-11-07
Device Manufacturer Date2017-06-09
Date Added to Maude2017-11-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactUSER SONIA IRIZARRY
Manufacturer Street311 ENTERPRISE DRIVE
Manufacturer CityPLAINSBORO NJ 08536
Manufacturer CountryUS
Manufacturer Postal08536
Manufacturer Phone6099362393
Manufacturer G1INTEGRA MICROFRANCE S.A.S.
Manufacturer StreetLE PAVILLON
Manufacturer CitySAINT AUBIN LE MONIAL 03160
Manufacturer CountryFR
Manufacturer Postal Code03160
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameSPECULUM MCO981D DEVEZE
Generic NamePFM11
Product CodeEPY
Date Received2017-11-15
Returned To Mfg2017-11-07
Catalog NumberMCO981D
Lot Number2172172
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerINTEGRA MICROFRANCE S.A.S.
Manufacturer AddressLE PAVILLON LE PAVILLON SAINT AUBIN LE MONIAL 03160 FR 03160


Patients

Patient NumberTreatmentOutcomeDate
10 2017-11-15