[93207102] The field service engineer investigated this incident on october 27, 2017. He inspected the table and tested it for drift. He was unable to verify the drift symptoms. After he removed the covers of the table, he observed some small hydraulic fluid leakage from the manifold area. This could potentially be the underlying cause of the drift. The field service engineer will be replacing the leaking hydraulic component(s) and performing functional testing to verify proper operation after the repair has been completed.
Patient Sequence No: 1, Text Type: N, H10


[93207103] The procedure had been successfully performed and mr imaging was completed. Hospital personnel were moving the magnet away from the table when they noticed contact between the head fixation device, which was attached to the table, and the bore of the magnet. This contact was not evident when the magnet was initially moved over the table and they assumed that the table drifted slightly downward. They lifted up on the back section of the table a small amount to eliminate the contact and then continued moving the magnet away from the table. There was no report of patient injury.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3010326005-2017-00011
MDR Report Key7034961
Date Received2017-11-15
Date of Report2017-12-19
Date of Event2017-10-12
Date Mfgr Received2017-10-12
Device Manufacturer Date2016-02-18
Date Added to Maude2017-11-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. TODD SPERLING
Manufacturer Street5101 SHADY OAK ROAD
Manufacturer CityMINNETONKA MN 553434100
Manufacturer CountryUS
Manufacturer Postal553434100
Manufacturer Phone7632036344
Manufacturer G1IMRIS - DEERFIELD IMAGING, INC.
Manufacturer Street5101 SHADY OAK ROAD
Manufacturer CityMINNETONKA MN 553434100
Manufacturer CountryUS
Manufacturer Postal Code553434100
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameIMRIS ORT300 OPERATING ROOM TABLE
Generic NameRADIOLOGIC TABLE
Product CodeKXJ
Date Received2017-11-15
Model NumberORT300
Catalog Number114093-000
OperatorNURSE
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerIMRIS - DEERFIELD IMAGING, INC.
Manufacturer Address5101 SHADY OAK ROAD MINNETONKA MN 553434100 US 553434100


Patients

Patient NumberTreatmentOutcomeDate
10 2017-11-15

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.