HOOK MCO205D 0.3MM 90DEG 165MM LFT ANG

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-11-16 for HOOK MCO205D 0.3MM 90DEG 165MM LFT ANG manufactured by Integra Microfrance S.a.s..

Event Text Entries

[93318100] Integra has completed their internal investigation on november 14, 2017. Results: evaluation of returned device; the hook is broken. The fragment was not recovered. A thorough observation of breakage area did not highlighted material or manufacturing defect. The etching was done with previous technology that was stopped on 2002. Dhr review; unable to review the applicable dhr as the lot no. Was unavailable. Complaints history; no adverse trend - first occurrence of this risk for this device. Conclusion: this device was manufactured before 2002 and no maintenance was performed by the integra service and repair. The age of the instrument and the lack of maintenance let reasonably conclude that this event is not attributable to integra.
Patient Sequence No: 1, Text Type: N, H10


[93318101] It was reported that the product's hook broke during surgery. The hook was on the device when the surgery began, but it was not there when starting the incision. It was impossible to determine if the hook was in the internal part of the patient's ear. The hook is ferromagnetic and they could not perform a mri.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2523190-2017-00143
MDR Report Key7038139
Date Received2017-11-16
Date of Report2017-10-26
Date Mfgr Received2017-10-26
Date Added to Maude2017-11-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactUSER SONIA IRIZARRY
Manufacturer Street311 ENTERPRISE DRIVE
Manufacturer CityPLAINSBORO NJ 08536
Manufacturer CountryUS
Manufacturer Postal08536
Manufacturer Phone6099362393
Manufacturer G1INTEGRA MICROFRANCE S.A.S.
Manufacturer StreetLE PAVILLON
Manufacturer CitySAINT AUBIN LE MONIAL 03160
Manufacturer CountryFR
Manufacturer Postal Code03160
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameHOOK MCO205D 0.3MM 90DEG 165MM LFT ANG
Generic NamePFM11
Product CodeJYL
Date Received2017-11-16
Returned To Mfg2017-11-13
Catalog NumberMCO205D
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerINTEGRA MICROFRANCE S.A.S.
Manufacturer AddressLE PAVILLON LE PAVILLON SAINT AUBIN LE MONIAL 03160 FR 03160


Patients

Patient NumberTreatmentOutcomeDate
10 2017-11-16

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.