MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-11-17 for COBAS 6000 C501 MODULE manufactured by Roche Diagnostics.
[93339794]
(b)(4). (b)(6).
Patient Sequence No: 1, Text Type: N, H10
[93339795]
The customer received questionable results for two patient samples. Of the data provided only the prea prealbumin results for one patient sample were a reportable malfunction. The initial result was 8 mg/dl. The sample was repeated and the results were 24 mg/dl and 9 mg/dl. No erroneous result was reported outside of the laboratory. There was no allegation of an adverse event. The reagent lot number was 274189. The expiration date was requested but was not provided. Based on analysis of the provided reaction curves, it was suspected the sample volume was too low.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1823260-2017-02656 |
MDR Report Key | 7042042 |
Report Source | FOREIGN,HEALTH PROFESSIONAL,U |
Date Received | 2017-11-17 |
Date of Report | 2018-01-29 |
Date of Event | 2017-10-31 |
Date Mfgr Received | 2017-10-31 |
Date Added to Maude | 2017-11-17 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 0 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | NA MICHAEL LESLIE |
Manufacturer Street | 9115 HAGUE ROAD NA |
Manufacturer City | INDIANAPOLIS IN 46250 |
Manufacturer Country | US |
Manufacturer Postal | 46250 |
Manufacturer Phone | 3175214343 |
Manufacturer G1 | HITACHI HIGH TECH CORP. |
Manufacturer Street | 882 ICHIGE HITACHINAKA NA |
Manufacturer City | IBARAKI 312-8504 |
Manufacturer Country | JA |
Manufacturer Postal Code | 312-8504 |
Single Use | 3 |
Previous Use Code | 3 |
Removal Correction Number | NA |
Event Type | 3 |
Type of Report | 3 |
Brand Name | COBAS 6000 C501 MODULE |
Generic Name | CLINICAL CHEMISTRY ANALYZER |
Product Code | CJW |
Date Received | 2017-11-17 |
Model Number | C501 |
Catalog Number | ASKU |
Lot Number | NA |
ID Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Eval'ed by Mfgr | R |
Device Sequence No | 0 |
Device Event Key | 0 |
Manufacturer | ROCHE DIAGNOSTICS |
Manufacturer Address | 9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457 |
Brand Name | COBAS 6000 C501 MODULE |
Generic Name | CLINICAL CHEMISTRY ANALYZER |
Product Code | JJE |
Date Received | 2017-11-17 |
Model Number | C501 |
Catalog Number | ASKU |
Lot Number | NA |
ID Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ROCHE DIAGNOSTICS |
Manufacturer Address | 9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2017-11-17 |