MODERMA FLEX UROSTOMY POUCH WITH CONVEX SKIN BARRIER 2969139

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,foreign,health pr report with the FDA on 2017-11-17 for MODERMA FLEX UROSTOMY POUCH WITH CONVEX SKIN BARRIER 2969139 manufactured by Hollister Incorporated.

Event Text Entries

[92360145] A device malfunction has not been alleged or determined. Additional information was received on november 16th from the clinical nurse who cared for the patient. The opinion of the treating nurse is that the presence of convexity against the patient's peristomal skin caused the pressure area. The use of convexity for patients with flush or receding stomas is the standard practice; however, the potential exists for pressure-related issues that should be monitored during its use to determine if the benefits of using the product outweigh the risk of effluent leakage.
Patient Sequence No: 1, Text Type: N, H10


[92360146] It was reported that the patient presented to the hospital on (b)(6) 2017 with a stage 2 pressure injury near his urinary stoma which was attributed to the hollister convex urostomy pouch. The patient was treated through (b)(6) 2017 with over 42 visits to the clinic. The pressure ulcer has healed. It was noted by the clinic that the patient did have a weight gain which is believed to have contributed to the pressure ulcer occurring.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1119193-2017-00036
MDR Report Key7042267
Report SourceDISTRIBUTOR,FOREIGN,HEALTH PR
Date Received2017-11-17
Date of Report2017-11-17
Date of Event2017-05-08
Report Date2017-09-06
Date Reported to Mfgr2017-09-06
Date Mfgr Received2017-09-20
Date Added to Maude2017-11-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR BENJAMIN LICHTENWALNER
Manufacturer Street2000 HOLLISTER DRIVE
Manufacturer CityLIBERTYVILLE IL 600483781
Manufacturer CountryUS
Manufacturer Postal600483781
Manufacturer Phone8479183497
Manufacturer G1HOLLISTER INCORPORATED
Manufacturer Street366 DRAFT AVENUE
Manufacturer CitySTUARTS DRAFT VA 244779998
Manufacturer CountryUS
Manufacturer Postal Code244779998
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMODERMA FLEX UROSTOMY POUCH WITH CONVEX SKIN BARRIER
Generic NameUROSTOMY POUCH
Product CodeEXG
Date Received2017-11-17
Model Number2969139
Catalog Number2969139
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerHOLLISTER INCORPORATED
Manufacturer Address366 DRAFT AVENUE STUARTS DRAFT VA 244779998 US 244779998


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2017-11-17

© 2026 FDA.report