RIGHT ANGLE AIRWAY PRONE 9600MR561

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06,07,08 report with the FDA on 2006-04-27 for RIGHT ANGLE AIRWAY PRONE 9600MR561 manufactured by Fisher & Paykel Healthcare, Ltd..

Event Text Entries

[456191] Fisher and paykel healthcare (fphc) received a report from cardinal health regarding an alleged fire. The patient had been on a fphc rt110 breathing circuit for about 24 hours. The humidifier was the fphc mr 370 and associated accessories, with mr290 humidification chamber. The patient, ventilator, humidifier and circuit were transported to the cath lab for a procedure. A respiratory therapist then went about his/her duties. During the procedure ( a period of about 1 hour) it was reported that the nursing staff silenced the mr730 alarms at least twice. The cause of the alarms is unknown. The respiratory therapist was called to respond to the cath lab stat and upon entering the area, immediately noted that the circuit was apparently melted and there allegedly were flames and sparks coming from the respiratory limb and a portion of the patient drape. The respiratory therapist immeidately assisted the patient suctioning and provided appropriate ventilation and oxygenation. There is no reported injury or concern for the patient at this time.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9611451-2006-00005
MDR Report Key704481
Report Source06,07,08
Date Received2006-04-27
Date of Report2006-03-28
Date of Event2006-03-27
Date Mfgr Received2006-03-31
Date Added to Maude2006-04-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactRENETA MONEY
Manufacturer Street15 MAURICE PAYKEL PLACE EAST TAMAKI, PO BOX 14348
Manufacturer CityPANMURE, AUCKLAND
Manufacturer CountryNZ
Manufacturer Phone5740100
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRIGHT ANGLE AIRWAY PRONE
Generic NameTEMPERATURE PROBE
Product CodeHCS
Date Received2006-04-27
Model Number9600MR561
Catalog Number9600MR561
Lot Number031015
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key693566
ManufacturerFISHER & PAYKEL HEALTHCARE, LTD.
Manufacturer Address* AUCKLAND NZ
Baseline Brand NameRIGHT ANGLE AIRWAY PRONE
Baseline Generic NameTEMPERATURE PROBE
Baseline Model No9600MR561
Baseline Catalog No9600MR561
Baseline IDNA


Patients

Patient NumberTreatmentOutcomeDate
10 2006-04-27

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.