CHROMID? MRSA SMART 413050

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-11-20 for CHROMID? MRSA SMART 413050 manufactured by Biomerieux Sa.

Event Text Entries

[93480605] A customer in poland notified biom? Rieux of a false positive associated with chromid? (b)(6). The customer reported receiving the (b)(6) smart medium growth of control strain s. Aureus (b)(6) atcc 43300 and two (2) (b)(6) strains on (b)(6) 2017. The medium was stored in accordance with the package insert at 2-8? C in the dark. The plates were exposed to light only for inoculation and reading. Repeat testing was completed on (b)(6) 2017 the test with the same strains on a different medium (b)(6) elite lot 1005970640 and received correct result - for control strain (b)(6) - growth, for (b)(6) strains - no growth. The (b)(6) strains were confirmed by various methods; disc diffusion method, vitek? 2, (b)(6) detection test for pbp2 protein with a negative result. There is no indication or report from the laboratory or physician that the discrepant result led to any adverse event related to a patient's state of health. An internal biom? Rieux investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2017-00377
MDR Report Key7044825
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-11-20
Date of Report2018-05-17
Date Mfgr Received2018-04-27
Device Manufacturer Date2017-09-08
Date Added to Maude2017-11-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. ELLEN WELTMER
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer G1BIOMERIEUX SA
Manufacturer Street5, RUE DES AQUEDUCS CRAPONNE, FR 69290
Manufacturer CountryFR
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCHROMID? MRSA SMART
Generic NameCHROMID? MRSA SMART
Product CodeJSO
Date Received2017-11-20
Catalog Number413050
Lot Number1005950400
Device Expiration Date2017-11-17
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX SA
Manufacturer Address5, RUE DES AQUEDUCS CRAPONNE, FR 69290 FR


Patients

Patient NumberTreatmentOutcomeDate
10 2017-11-20

© 2024 FDA.report