ROTO-CAM SUPERSCS MON 3.5X24 630202

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,foreign report with the FDA on 2017-11-20 for ROTO-CAM SUPERSCS MON 3.5X24 630202 manufactured by Integra York, Pa Inc..

Event Text Entries

[93346954] 11/8/2017 integra investigation completed. Method: failure analysis, device history evaluation. Results: failure analysis - roto-cam scissors returned in used condition, not showing any unusual markings. There is a yellow tape marking and the hinge at tip is broken. This may happen if too much pressure is used to the tip, when in use. The complaint report is confirmed; damaged worn. Device history evaluation: device history record reviewed and no anomalies found for reported incident.. Conclusion: the root cause has not been identified as a workmanship or material deficiency.
Patient Sequence No: 1, Text Type: N, H10


[93346955] Customer initially reports: during the surgery, the customer found that the hinge of the jaw was damaged and came off, though it was the second time use of the device. They are afraid that the parts of the device still remained in patient? S abdomen. (b)(6) 2017 customer reports laparoscopic cholecystectomy being performed. They did x-ray to confirm whether the device is xray transparent or not. No indication that any part was visualized. No harm identified at this time.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2523190-2017-00130
MDR Report Key7045194
Report SourceDISTRIBUTOR,FOREIGN
Date Received2017-11-20
Date of Report2017-10-27
Date of Event2017-10-18
Date Mfgr Received2017-10-27
Device Manufacturer Date2016-06-01
Date Added to Maude2017-11-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactUSER SANDRA LEE
Manufacturer Street311 ENTERPRISE DRIVE
Manufacturer CityPLAINSBORO NJ 08536
Manufacturer CountryUS
Manufacturer Postal08536
Manufacturer Phone6099362393
Manufacturer G1INTEGRA YORK, PA INC.
Manufacturer Street589 DAVIES DRIVE
Manufacturer CityYORK PA 17402
Manufacturer CountryUS
Manufacturer Postal Code17402
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameROTO-CAM SUPERSCS MON 3.5X24
Generic Name3.5MM ENDOSCOPIC INSTRUMENTATION
Product CodeKNF
Date Received2017-11-20
Returned To Mfg2017-11-03
Catalog Number630202
Lot Number100228-1606
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerINTEGRA YORK, PA INC.
Manufacturer Address589 DAVIES DRIVE 589 DAVIES DRIVE YORK PA 17402 US 17402


Patients

Patient NumberTreatmentOutcomeDate
10 2017-11-20

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.