ELECTROSURGICAL PENCIL

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2017-11-20 for ELECTROSURGICAL PENCIL manufactured by Megadyne Medical Products.

Event Text Entries

[93368703] A review of the device history record shows this product was originally manufactured to specifications. A review of complaint history was conducted from the last two years and indicated no other reports of this same issue have been reported which indicated a malfunction. Also search of complaint history for reported lot number resulted in no other complaints issued. During initial evaluation, the device was plugged into a esu and continuous electrical connection was observed. The device has been forwarded to megadyne's supplier and investigation is on going to determine cause.
Patient Sequence No: 1, Text Type: N, H10


[93368704] Bovie pencil was plugged in and automatically turned on. Unplugged and button was not pressed, plugged in again and pencil automatically turned on again. Bovie was replaced with a different bovie with no issues. No patient injury.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1721194-2017-00009
MDR Report Key7045684
Report SourceHEALTH PROFESSIONAL
Date Received2017-11-20
Date of Report2017-11-27
Date Mfgr Received2017-10-24
Device Manufacturer Date2017-08-17
Date Added to Maude2017-11-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS STACEY CASTANEDA
Manufacturer Street11506 SOUTH STATE STREET
Manufacturer CityDRAPER UT 84020
Manufacturer CountryUS
Manufacturer Postal84020
Manufacturer Phone8007476110
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameELECTROSURGICAL PENCIL
Generic NameELECTROSURGICAL PENCIL
Product CodeJOS
Date Received2017-11-20
Returned To Mfg2017-10-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEGADYNE MEDICAL PRODUCTS
Manufacturer Address11506 SOUTH STATE STREET DRAPER UT 84118 US 84118


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-11-20

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.