LIGACLIP*ENDO MED/LG APPLIER EL314

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-11-22 for LIGACLIP*ENDO MED/LG APPLIER EL314 manufactured by Ethicon Endo-surgery, Llc..

Event Text Entries

[93584284] (b)(4). Device analysis: the el314 device was returned for analysis and upon inspection the jaws were found to be in a yielded condition. No functional test was performed due to the condition of the applier. Possible causes for the condition found may be if the device is closed over an existing hard object or clip placing stress on the jaws causing them to distort or yield and not return to their original dimensions/position or excessive application of torque to the jaws when positioning the device on a vessel. It should be noted that as part of our quality process, each device is visually inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment. The batch history records were reviewed and certed by external manufacturing that the manufacturing criteria was met prior to the release of the equipment.?
Patient Sequence No: 1, Text Type: N, H10


[93584285] It was reported that the device was found to not be holding clips. There was no patient involvement.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005075853-2017-06275
MDR Report Key7052444
Date Received2017-11-22
Date of Report2017-10-31
Date of Event2017-08-28
Date Mfgr Received2017-10-31
Device Manufacturer Date2005-11-30
Date Added to Maude2017-11-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMILTON GARRETT
Manufacturer Street475 CALLE C
Manufacturer CityGUAYNABO 00969
Manufacturer Postal00969
Manufacturer Phone5133378865
Manufacturer G1ETHICON ENDO-SURGERY, LLC.
Manufacturer Street475 CALLE C
Manufacturer CityGUAYNABO 00969
Manufacturer Postal Code00969
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameLIGACLIP*ENDO MED/LG APPLIER
Generic NameLIGACLIP
Product CodeHBT
Date Received2017-11-22
Returned To Mfg2017-10-06
Catalog NumberEL314
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerETHICON ENDO-SURGERY, LLC.
Manufacturer Address475 CALLE C GUAYNABO 00969 00969


Patients

Patient NumberTreatmentOutcomeDate
10 2017-11-22

© 2024 FDA.report