MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2017-11-22 for ORS-320 SLUSH WARMER DISC DRAPE manufactured by Microtek Dominicana, S.a..
[93300168]
No samples were returned for this investigation and a lot number was not provided to perform a dhr review. Since no lot number or sample was provided the non conformity could not be confirmed. It was not determined if this issue was related to a personnel, process or material issue. Melts, though uncommon, can occur when the warmer is not draped properly, not turned off by using the power button, and/or there is insufficient fluid in the warmer basin, contrary to the operations manual, product labeling, product insert and in-service presentations. If additional information becomes available, a follow up report will be submitted.
Patient Sequence No: 1, Text Type: N, H10
[93300169]
End user reported " surgery completed in 501, during clean up after case, it was noted that there was irrigation fluid in the bottom of the fluid warmer, underneath the drape. This would indicate there was a hole in the fluid warmer drape used during the surgical procedure. There is no way to know when the hole occurred. Unable to pull lot number from drape, package had already been discarded. " no patient injury or treatment was reported for the incident.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 8043817-2017-00032 |
MDR Report Key | 7052795 |
Report Source | COMPANY REPRESENTATIVE,HEALTH |
Date Received | 2017-11-22 |
Date of Report | 2017-10-24 |
Date of Event | 2017-03-30 |
Date Mfgr Received | 2017-10-24 |
Date Added to Maude | 2017-11-22 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MRS JENNIFER MRKVICKA |
Manufacturer Street | 1 ECOLAB PLACE |
Manufacturer City | ST. PAUL MN 55102 |
Manufacturer Country | US |
Manufacturer Postal | 55102 |
Manufacturer Phone | 6512502237 |
Manufacturer G1 | MICROTEK DOMINICANA |
Manufacturer Street | ZONA FRANCA NO. 2 |
Manufacturer City | LA ROMANA, |
Manufacturer Country | DR |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | ORS-320 SLUSH WARMER DISC DRAPE |
Generic Name | FLUID WARMING AND SLUSH DRAPES |
Product Code | LHC |
Date Received | 2017-11-22 |
Model Number | ORS-320 |
Lot Number | UNKNOWN |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | MICROTEK DOMINICANA, S.A. |
Manufacturer Address | ZONA FRANCA NO. 2 LA ROMANA, DR |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2017-11-22 |