OPTICHAMBER DIAMOND 1098214

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-11-23 for OPTICHAMBER DIAMOND 1098214 manufactured by Respironics Respiratory Drug Delivery (uk) Ltd.

Event Text Entries

[92839671] The optichamber diamond is available with three mask sizes - small, medium and large. Instructions for use supplied with the device states 'recommended patient population: optichamber diamond with small litetouch mask: 0 to 18 months. Optichamber diamond with medium mask: 1 to 5 years. Optichamber diamond with large litetouch mask: 5 years +. ' as the device is restricted to sale by or on the order of a physician or licensed healthcare professional, it is expected that the patient be assessed for adequate fit prior to use, and appropriate mask selected. Device is not life sustaining/ supporting. No other complaints received relating to recommended sizing issues.
Patient Sequence No: 1, Text Type: N, H10


[92839672] Patient caregiver reported that the (b)(6) patient was admitted to hospital overnight due to use of an optichamber diamond valved holding chamber device with a small litetouch mask that was too small for the patient, and therefore patient did not get the medication needed. Patient was discharged with a larger mask. Caregiver states her daughter is doing fine now.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9681154-2017-00020
MDR Report Key7057130
Date Received2017-11-23
Date of Report2017-11-09
Date of Event2017-11-08
Date Mfgr Received2017-11-09
Date Added to Maude2017-11-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR ALESSANDRO AGOSTI
Manufacturer StreetCHICHESTER BUSINESS PARK CITY TANGMERE
Manufacturer CityCHICHESTER, PO202FT
Manufacturer CountryUK
Manufacturer PostalPO202FT
Manufacturer Phone8704231549
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameOPTICHAMBER DIAMOND
Generic NameHOLDING CHAMBERS, DIRECT PATIENT INTERFACE
Product CodeNVP
Date Received2017-11-23
Model Number1098214
Catalog Number1098214
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerRESPIRONICS RESPIRATORY DRUG DELIVERY (UK) LTD
Manufacturer AddressCHICHESTER BUSINESS PARK CITY FIELDS WAY, TANGMERE CHICHESTER, PO202FT UK PO202FT


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization 2017-11-23