MONOMAX VIOLET 1 (4) 90CM HR40S (M) B0041098

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2017-11-27 for MONOMAX VIOLET 1 (4) 90CM HR40S (M) B0041098 manufactured by B.braun Surgical Sa.

Event Text Entries

[93228321] (b)(4). Manufacturing site evaluation: samples received: 2 unopened pouches. Analysis and results: there are no previous complaints of this code batch of which we manufactured and distributed (b)(4) units in the market. There are no units in the warehouse. Tightness test to the closed samples received has been performed and the units are tight. Tested the knot pull tensile strength of the closed samples received and the results fulfill the oem requirements. Reviewed the batch manufacturing record, this product had a normal process and the results during the process fulfill the requirements. Final conclusion: although the results of the samples received fulfill the oem specifications, note of this incident is taken in order to assess if new or additional actions are needed. Corrective/preventive actions: according to our internal procedures, there is no need to establish corrective or preventive actions. Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
Patient Sequence No: 1, Text Type: N, H10


[93228322] Country of complaint: (b)(6). Suture breakage while pulling the thread.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3003639970-2017-00564
MDR Report Key7061904
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2017-11-27
Date of Report2017-11-27
Date of Event2017-10-03
Date Facility Aware2017-11-10
Date Mfgr Received2017-10-31
Date Added to Maude2017-11-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. NICOLE BROYLES
Manufacturer Street615 LAMBERT POINTE DRIVE
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer G1B.BRAUN SURGICAL SA
Manufacturer Street121 CARRETERA DE TERRASSA
Manufacturer CityRUBI, 08191
Manufacturer CountrySP
Manufacturer Postal Code08191
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMONOMAX VIOLET 1 (4) 90CM HR40S (M)
Generic NameSUTURES
Product CodeNWJ
Date Received2017-11-27
Returned To Mfg2017-11-06
Model NumberB0041098
Catalog NumberB0041098
Lot Number115392
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerB.BRAUN SURGICAL SA
Manufacturer Address121 CARRETERA DE TERRASSA RUBI, 08191 SP 08191


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-11-27

© 2024 FDA.report