RESQPUMP

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2017-11-29 for RESQPUMP manufactured by .

Event Text Entries

[93703479] There is damage to the key in the lower clamshell that prevents the stem from turning, that could have occurred when the stem came off and another compression came down, binding up the guide. There is not similar damage on the mating notch in the stem. There is white plastic shavings inside the socket on the suction cup that the stem inserts into. There is also some of the same inside the stem, likely from rubbing on the spring. The original pink loctite is visible at the base of the bevel on the screw head. There is also green thread locker on about the last 3/8" of threads. That's not factory issue, so applied afterwards. The threads aren't stripped out, and there's not much evidence of wear,,, brass shavings or corrosion. The screw threads into the brass spring end easily, there is some play, not excessive. No evidence of tampering or damage was present on the inside of the clamshell.
Patient Sequence No: 1, Text Type: N, H10


[93703480] During a cardiac arrest, yesterday, (b)(6) 2017 we utilized the resqpump on a female patient. During use (a 20+ minute on scene evolution) the phillips head screw that is vertically installed in the white plunger section backed out and caused the pump to come apart.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3003477173-2017-00009
MDR Report Key7069048
Report SourceHEALTH PROFESSIONAL
Date Received2017-11-29
Date of Report2017-11-29
Date of Event2017-10-26
Date Mfgr Received2017-11-06
Date Added to Maude2017-11-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS DIANE HOWELL
Manufacturer Street1905 COUNTY ROAD C WEST
Manufacturer CityROSEVILLE MN 55113
Manufacturer CountryUS
Manufacturer Postal55113
Manufacturer Phone6514035600
Manufacturer G1ADVANCED CIRCULATORY SYSTEMS
Manufacturer Street1905 COUNTY ROAD C WEST
Manufacturer CityROSEVILLE MI 55113
Manufacturer CountryUS
Manufacturer Postal Code55113
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRESQPUMP
Generic NameRESQPUMP
Product CodePIZ
Date Received2017-11-29
Device Availability*
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0


Patients

Patient NumberTreatmentOutcomeDate
10 2017-11-29

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.