CHROMID? CPS? ELITE AGAR 416172

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-11-30 for CHROMID? CPS? ELITE AGAR 416172 manufactured by Biom?rieux Sa.

Event Text Entries

[94220702] A customer in (b)(6) notified biom? Rieux of false negative results for two escherichia coli strains. The two strains were growing in white colonies, rather than the characteristically pink colonies, on chromid? Cps? Elite media plate, lot 1005953180. The identification of the strains were confirmed as escherichia coli by maldi-tof. There is no indication or report from the laboratory that the discrepant result led to any adverse event related to any patient's state of health. A biom? Rieux internal investigation has been initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2017-00387
MDR Report Key7071461
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-11-30
Date of Report2017-12-22
Date Mfgr Received2017-12-04
Device Manufacturer Date2017-09-08
Date Added to Maude2017-11-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRADIOLOGIC TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. ELLEN WELTMER
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147317301
Manufacturer G1BIOM
Manufacturer Street5 RUE DES AQUEDUCS
Manufacturer CityCRAPONNE, 69290
Manufacturer CountryFR
Manufacturer Postal Code69290
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCHROMID? CPS? ELITE AGAR
Generic NameCHROMID? CPS? ELITE AGAR
Product CodeJXA
Date Received2017-11-30
Catalog Number416172
Lot Number1005953180
Device Expiration Date2018-01-04
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBIOM?RIEUX SA
Manufacturer Address5 RUE DES AQUEDUCS CRAPONNE, 69290 FR 69290


Patients

Patient NumberTreatmentOutcomeDate
10 2017-11-30

© 2026 FDA.report