NAVICARE NURSE CALL NNC 3.6.600

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-11-30 for NAVICARE NURSE CALL NNC 3.6.600 manufactured by Hill-rom Cary.

Event Text Entries

[94279697] Hill-rom investigation determined the patient room was offline at the time of the reported event. The appropriate system alerts were present on the system in the room and at the nurse? S station indicating the room was offline. A review of the log files indicates a nurse was present in the room for 1 1/2 minutes prior to pulling the code blue lever with two additional staff members entering the room prior to pulling the code blue lever. The staff remained in the room at the time the device did not ring remotely at the primary nurse station. Hill-rom requested the return of the room control board for investigation. The user facility declined to release the room control board. Investigation to date indicates the control board for the room as the potential cause of the room dropping off line. If additional information becomes available during our investigation, a follow up report will be filed.
Patient Sequence No: 1, Text Type: N, H10


[94279698] Hill-rom received a report from the account stating the code blue lever was pulled but that the call was delayed in ringing at the primary nurse station.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2027454-2017-00001
MDR Report Key7073275
Date Received2017-11-30
Date of Report2017-10-31
Date of Event2017-10-22
Date Mfgr Received2017-10-31
Device Manufacturer Date2016-12-01
Date Added to Maude2017-11-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR GEOFF GORHAM
Manufacturer Street1225 CRESCENT GREEN DR., SUITE 300
Manufacturer CityCARY NC 27518
Manufacturer CountryUS
Manufacturer Postal27518
Manufacturer G1HILL-ROM
Manufacturer Street1225 CRESCENT GREEN DR., SUITE
Manufacturer CityCARY NC 27518
Manufacturer CountryUS
Manufacturer Postal Code27518
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameNAVICARE NURSE CALL
Generic NameNAVICARE NURSE CALL
Product CodeILQ
Date Received2017-11-30
Model NumberNNC 3.6.600
OperatorNURSE
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerHILL-ROM CARY
Manufacturer Address1225 CRESCENT GREEN DR., SUITE 300 CARY NC 27518 US 27518


Patients

Patient NumberTreatmentOutcomeDate
101. Death 2017-11-30

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.