VIDAS? LYME IGM 30319

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-12-01 for VIDAS? LYME IGM 30319 manufactured by Biomerieux, Sa.

Event Text Entries

[94570736] A customer in (b)(6) reported to biom? Rieux that they have observed false negative results when using the products vidas? Lyme igm (ref. 30319) lot 1005810050 and vidas? Lyme igg (ref. 30320) lot 1005816250. The patient is a (b)(6) male with complaints of joint pain. The patient did not remember having a tick bite and erythema migrans was not observed. On (b)(6) 2017, the customer tested the patient's sample on both vidas? Lyme igm and vidas? Lyme igg. The results obtained were negative with the following values:? Lyme igg: 0. 10,? Lyme igm: 0. 02. The patient's sample had been tested with western blot technique (borrelia euroline-wb igg). The results were positive with 3 bands for igm and 4 bands for igg. The customer reported that patient results were affected and wrong results were reported to a physician. The customer stated there was no delay in reporting results. There is no indication or report from the laboratory that the discrepant result led to any adverse event related to any patient's state of health. A biom? Rieux internal investigation has been initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2017-00388
MDR Report Key7074792
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-12-01
Date of Report2018-01-31
Date Mfgr Received2018-01-04
Device Manufacturer Date2017-06-30
Date Added to Maude2017-12-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. ELLEN WELTMER
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147317301
Manufacturer G1BIOMERIEUX, SA
Manufacturer Street376 CHEMIN DE L'ORME
Manufacturer CityMARCY L'ETOILE, 69280
Manufacturer CountryFR
Manufacturer Postal Code69280
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVIDAS? LYME IGM
Generic NameVIDAS? LYME IGM
Product CodeLSR
Date Received2017-12-01
Catalog Number30319
Lot Number1005810050
Device Expiration Date2018-06-14
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX, SA
Manufacturer Address376 CHEMIN DE L'ORME MARCY L'ETOILE, 69280 FR 69280


Patients

Patient NumberTreatmentOutcomeDate
10 2017-12-01

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.