NEVADA ELECTRIC 4C 32X4.5 INCHES 110V SPA TABLE 414921

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-12-04 for NEVADA ELECTRIC 4C 32X4.5 INCHES 110V SPA TABLE 414921 manufactured by Silhouet Tone Corporation.

Event Text Entries

[94138638] Overheating of the heated cushion due to interaction with memory foam. No client on the spa table at the moment it happened. The unit has been repaired on site. The product is now manufactured with another foam.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005545844-2017-01605
MDR Report Key7078911
Date Received2017-12-04
Date of Report2017-10-16
Date of Event2016-10-20
Date Mfgr Received2016-10-20
Device Manufacturer Date2014-04-08
Date Added to Maude2017-12-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS GENEVIEVE ST-JEAN
Manufacturer Street2185 MICHELIN
Manufacturer CityLAVAL, QU H7L5B8
Manufacturer CountryCA
Manufacturer PostalH7L5B8
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameNEVADA ELECTRIC 4C 32X4.5 INCHES 110V SPA TABLE
Generic NameSPA TABLE
Product CodeLGX
Date Received2017-12-04
Model Number414921
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSILHOUET TONE CORPORATION
Manufacturer Address2185 MICHELIN LAVAL, QU?BEC H7L5B8 CA H7L5B8


Patients

Patient NumberTreatmentOutcomeDate
10 2017-12-04

© 2026 FDA.report