ACCURUN 1 MULTI-MARKER POSITIVE CONTROL SERIES 4400 A001-4408-P

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2006-04-28 for ACCURUN 1 MULTI-MARKER POSITIVE CONTROL SERIES 4400 A001-4408-P manufactured by Bbi Diagnostics.

Event Text Entries

[456600] Technologist report eye splash with product. She experienced difficulty removing the push and pull cap from a vial of the product (the product is manufactured and shipped with a screw cap, but the laboratory replaced the screw cap with a push and pull cap). When she applied forceto the cap, a drop of product on the cap splashed into her left eye. Technologist was not wearing protective eyewear and was wearing contact lenses. She removed her contact lens and flushed her eye with water for approximately 10 minutes. Technologist was referred to the hospital's outpaient infectious disease service for follow-up medical evaluation, as the product contains chemically treated hiv and hepatitis positive raw materials. As a precaution, the technologist received anti-hiv prohylaxis and a hepatitis b vaccine, although the risk of infection was felt to be low.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1220394-2006-00001
MDR Report Key707895
Report Source05
Date Received2006-04-28
Date of Report2006-04-28
Date of Event2006-04-11
Date Mfgr Received2006-04-11
Device Manufacturer Date2005-09-01
Date Added to Maude2006-05-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactDEBRA PETIT
Manufacturer Street375 WEST STREET
Manufacturer CityWEST BRIDGEWATER MA 02379
Manufacturer CountryUS
Manufacturer Postal02379
Manufacturer Phone5085801900
Manufacturer G1BBI DIAGNOSTICS, A SERACARE COMPANY
Manufacturer Street375 WEST STREET
Manufacturer CityWEST BRIDGEWATER MA 02379
Manufacturer CountryUS
Manufacturer Postal Code02379
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameACCURUN 1 MULTI-MARKER POSITIVE CONTROL
Generic Name82MYZ
Product CodeMJX
Date Received2006-04-28
Model NumberSERIES 4400
Catalog NumberA001-4408-P
Lot Number109744
ID Number*
Device Expiration Date2008-09-23
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key696946
ManufacturerBBI DIAGNOSTICS
Manufacturer Address* WEST BRIDGEWATER MA * US
Baseline Brand NameACCURUN1 MULTI-MARKER POSITIVE CONTROL
Baseline Generic Name82MYZ
Baseline Model NoSERIES 4400
Baseline Catalog NoA00-4408 P
Baseline ID*


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2006-04-28