VIDAS? LYME IGM 30319

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-12-05 for VIDAS? LYME IGM 30319 manufactured by Biomerieux Sa.

Event Text Entries

[93735482] A customer in the (b)(6) notified biom? Rieux of discrepant results associated with vidas? Lyme igm (reference 30319). The customer reported having a recent patient sample that flagged igg positive and igm negative on the initial screening with vidas?. The sample was sent to a reference laboratory and the result was positive for both igg and igm. The reference laboratory borrelia igm lineblot indicated antibody to osp17 antigen to be detected. Patient treatment was delayed resulting in further issues. A biom? Rieux investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2017-00392
MDR Report Key7083102
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-12-05
Date of Report2018-02-16
Date Mfgr Received2018-01-23
Device Manufacturer Date2017-01-27
Date Added to Maude2017-12-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. ELLEN WELTMER
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityST. LOUIS MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer G1BIOMERIEUX SA
Manufacturer StreetCHEMIN DE L ORME MARCY L ETOILE, RHONE 69280
Manufacturer CountryFR
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVIDAS? LYME IGM
Generic NameVIDAS? LYME IGM
Product CodeLSR
Date Received2017-12-05
Catalog Number30319
Lot Number1005460450
Device Expiration Date2017-12-12
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX SA
Manufacturer AddressCHEMIN DE L ORME MARCY L ETOILE, RHONE 69280 FR


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2017-12-05

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.