TRIAGE D-DIMER PANEL 98100EU

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer,foreign,health profe report with the FDA on 2017-12-05 for TRIAGE D-DIMER PANEL 98100EU manufactured by Quidel Cardiovascular Inc..

Event Text Entries

[93804403] Customer's complaint was not replicated with in-house testing of retain lot w62838rb. No issues with analyse recovery were observed and the product performed as expected. Manufacturing batch records were reviewed and found that the lot met release specifications. Based on the information available, there is no indication of a product deficiency and no corrective action is required.
Patient Sequence No: 1, Text Type: N, H10


[93804404] On (b)(6) 2017: patient came to physician due to pain in left leg. Triage d-dimer result was 1240ng/ml. The patient was treated with blood thinner, lixiana and compression. Customer is not questioning this data, only provded as historical information. On (b)(6) 2017: patient came back to physician with pain in left leg. Triage d-dimer result was 1050ng/ml. Due to this result, patient was transferred to angiographgy institue for verification. A ct scan was performed. No deep vein thrombosis was detected.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2024674-2017-00004
MDR Report Key7085393
Report SourceCONSUMER,FOREIGN,HEALTH PROFE
Date Received2017-12-05
Date of Report2017-12-05
Date of Event2017-10-23
Date Added to Maude2017-12-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJESSICA PERROTTE
Manufacturer Street9975 SUMMERS RIDGE ROAD
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8588053573
Manufacturer G1QUIDEL CARDIOVASCULAR INC.
Manufacturer Street9975 SUMMERS RIDGE ROAD
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal Code92121
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRIAGE D-DIMER PANEL
Generic NameD-DIMER: FIBRIN SPLIT PRODUCTS
Product CodeGHH
Date Received2017-12-05
Model Number98100EU
Lot NumberW62838RB
Device Expiration Date2017-11-12
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerQUIDEL CARDIOVASCULAR INC.
Manufacturer Address9975 SUMMERS RIDGE ROAD SAN DIEGO CA 92121 US 92121


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-12-05