HELIUM CYLINDER UHP6K

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-12-05 for HELIUM CYLINDER UHP6K manufactured by Airgas Healthcare.

Event Text Entries

[94161300] A tumor cryoablation device from healthtronics was tested in the (b)(6) hospital operating room just prior to use. Patient was anesthetized and prepared for procedure. The device uses argon gas to freeze the sterile probe, helium to warm it back up. After freezing, it was found that the procedure to warm the probe did not work. The procedure was canceled with no adverse effect on patient. The mfg. Rep returned the next day to troubleshoot the problem. It was confirmed that 3 helium tanks in the same lot did not warm the probe but instead, cooled it to around minus 7 degrees c. Testing with a helium tank from another lot number confirmed that the device operated properly.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5073754
MDR Report Key7086701
Date Received2017-12-05
Date of Report2017-12-04
Date of Event2017-10-24
Date Added to Maude2017-12-06
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Sequence Number: 1

Brand NameHELIUM CYLINDER UHP6K
Generic NameHELIUM CYLINDER UHP6K
Product CodeBSE
Date Received2017-12-05
Returned To Mfg2017-11-29
Lot Number166-400878318-1
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrI
Device Sequence No1
Device Event Key0
ManufacturerAIRGAS HEALTHCARE
Manufacturer AddressRIVERSIDE CA 92507 US 92507

Device Sequence Number: 2

Brand NameHELIUM CYLINDER UHP6K
Generic NameHELIUM CYLINDER UHP6K
Product CodeBSE
Date Received2017-12-05
Returned To Mfg2017-11-29
Lot Number166-400878318-1
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrI
Device Sequence No2
Device Event Key0
ManufacturerAIRGAS HEALTHCARE
Manufacturer AddressRIVERSIDE CA 92507 US 92507

Device Sequence Number: 3

Brand NameHELIUM CYLINDER UHP6K
Generic NameHELIUM CYLINDER UHP6K
Product CodeBSE
Date Received2017-12-05
Returned To Mfg2017-11-29
Lot Number166-400878318-1
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrI
Device Sequence No3
Device Event Key0
ManufacturerAIRGAS HEALTHCARE
Manufacturer AddressRIVERSIDE CA 92507 US 92507


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2017-12-05

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.