CELLEBRITY? M00516011 1601

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2017-12-06 for CELLEBRITY? M00516011 1601 manufactured by Boston Scientific - Spencer.

Event Text Entries

[94833401] Patient's exact age is unknown; however it was reported that the patient was over the age of 18. (b)(4). Investigation results: visual evaluation of the returned celebrity cytology brush revealed that the catheter was kinked and torn near to the handle. Further evaluation noted that the pull wire was found kinked and broken with evidence of bending. Functional analysis showed that the brush failed to extend due to the condition of the returned device. Based on the evaluation of the returned device, these failures are likely due to anatomical or procedural factors encountered during the procedure which limited the performance of the device. Therefore, the most probable root cause is operational context. A review of the device history record (dhr) was performed and no deviations were found.
Patient Sequence No: 1, Text Type: N, H10


[94833402] It was reported to boston scientific corporation that a celebrity cytology brush was used in the lungs during a bronchoscopy procedure. According to the complainant, during the procedure, the catheter was kinked and the brush failed to extend and retract. The procedure was completed with another celebrity cytology brush. The patient's condition at the conclusion of the procedure was reported to be stable. This event has been deemed a reportable event based on the investigation results; wire broke.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005099803-2017-03503
MDR Report Key7087270
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2017-12-06
Date of Report2017-11-10
Date Mfgr Received2017-11-10
Device Manufacturer Date2017-03-22
Date Added to Maude2017-12-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactEMP. NANCY CUTINO
Manufacturer Street100 BOSTON SCIENTIFIC WAY
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5086834000
Manufacturer G1BOSTON SCIENTIFIC - SPENCER
Manufacturer Street780 BROOKSIDE DRIVE
Manufacturer CitySPENCER IN 47460
Manufacturer CountryUS
Manufacturer Postal Code47460
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCELLEBRITY?
Generic NameBRUSH, BIOPSY, BRONCHOSCOPE (NON-RIGID)
Product CodeBTG
Date Received2017-12-06
Returned To Mfg2017-10-27
Model NumberM00516011
Catalog Number1601
Lot Number20420907
Device Expiration Date2019-03-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC - SPENCER
Manufacturer Address780 BROOKSIDE DRIVE SPENCER IN 47460 US 47460


Patients

Patient NumberTreatmentOutcomeDate
10 2017-12-06

© 2026 FDA.report