VITEK? 2 AST-YS07 TEST KIT 414967

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-12-07 for VITEK? 2 AST-YS07 TEST KIT 414967 manufactured by Biomerieux, Inc.

Event Text Entries

[95764192] A customer from (b)(6) reported to biom? Rieux a false susceptible fluconazole result for a candida albicans strain from a patient blood culture sample, in association with the vitek? 2 ast-ys07 test kit. The customer reported the initial fluconazole result with the ast-ys07 card was susceptible (mic =4 mg/l). The strain was sent to a reference lab and the result was resistant (mic>= 64? G/ml) for fluconazole. There was no retest performed on the ast-ys07 card. The customer stated the false susceptible result was reported to the physician, was then corrected following lab reports from the reference lab. There is no indication or report from the hospital or treating physician to biom? Rieux that the discrepant result led to any adverse event related to the patient's state of health. A biom? Rieux internal investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1950204-2017-00435
MDR Report Key7096329
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-12-07
Date of Report2018-05-14
Date Mfgr Received2018-04-17
Device Manufacturer Date2017-03-06
Date Added to Maude2017-12-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. ELLEN WELTMER
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityST. LOUIS MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147317301
Manufacturer G1BIOMERIEUX, INC
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityST. LOUIS MO 63042
Manufacturer CountryUS
Manufacturer Postal Code63042
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVITEK? 2 AST-YS07 TEST KIT
Generic NameVITEK? 2 AST-YS07 TEST KIT
Product CodeNGZ
Date Received2017-12-07
Catalog Number414967
Lot Number2870283203
ID Number03573026426873
Device Expiration Date2018-09-05
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX, INC
Manufacturer Address595 ANGLUM ROAD ST. LOUIS MO 63042 US 63042


Patients

Patient NumberTreatmentOutcomeDate
10 2017-12-07

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.