LEICA RM2255 1492255UL01

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2017-12-08 for LEICA RM2255 1492255UL01 manufactured by Leica Biosystems Nussloch Gmbh.

Event Text Entries

[94223035] The investigation revealed the following: this incident is user related because the user disregarded the safety instructions. The customer forgot to engage the lock and did not use the safety cover for the blade. The injury was not an instrument malfunction. A safety training was performed with the customer to avoid this injury in future.
Patient Sequence No: 1, Text Type: N, H10


[94223036] On (b)(6) 2017, leica biosystems received a complaint that a customer injured his finger during the usage of the microtome. Medical treatment was necessary.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8010478-2017-00008
MDR Report Key7099263
Report SourceHEALTH PROFESSIONAL
Date Received2017-12-08
Date of Report2017-11-10
Date of Event2017-10-20
Date Mfgr Received2017-11-13
Device Manufacturer Date2016-12-16
Date Added to Maude2017-12-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. ROBERT GROPP
Manufacturer StreetHEIDELBERGERSTR. 17-19
Manufacturer CityNUSSLOCH, 69226
Manufacturer CountryGM
Manufacturer Postal69226
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLEICA RM2255
Generic NameMICROTOM, PRODUCT CODE: IDO
Product CodeIDO
Date Received2017-12-08
Model Number1492255UL01
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerLEICA BIOSYSTEMS NUSSLOCH GMBH
Manufacturer AddressHEIDELBERGERSTR. 17 NUSSLOCH, 69226 GM 69226


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-12-08

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.