MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-12-11 for VITEK? 2 AST-YS07 TEST KIT 414967 manufactured by Biomerieux, Inc..
[95746868]
A customer in (b)(4) reported the occurrence of false susceptible flucytosine result (mic = 4? G/ml) for candida norvegensis eeq biologie proficiency sample in association with the vitek? 2 ast-ys07 test kit. The expected result was "resistant"; no expected mic was provided for the proficiency sample. The vitek? 2 ast-ys07 flucytosine result mic = 4 is susceptible (s) using eucast breakpoints, but resistant (r) using clsi breakpoints. There is no indication or report from the laboratory that the discrepant result led to any adverse event related to any patient's state of health. There was no patient directly associated with the proficiency sample. Culture submittal has been requested by biom? Rieux for internal investigation. Biom? Rieux investigation has been initiated.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1950204-2017-00443 |
MDR Report Key | 7104475 |
Report Source | FOREIGN,HEALTH PROFESSIONAL,U |
Date Received | 2017-12-11 |
Date of Report | 2018-05-10 |
Date Mfgr Received | 2018-04-17 |
Device Manufacturer Date | 2017-01-17 |
Date Added to Maude | 2017-12-11 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MRS. ELLEN WELTMER |
Manufacturer Street | 595 ANGLUM ROAD |
Manufacturer City | HAZELWOOD MO 63042 |
Manufacturer Country | US |
Manufacturer Postal | 63042 |
Manufacturer Phone | 3147317301 |
Manufacturer G1 | BIOMERIEUX, INC. |
Manufacturer Street | 595 ANGLUM ROAD |
Manufacturer City | HAZELWOOD MO 63042 |
Manufacturer Country | US |
Manufacturer Postal Code | 63042 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | VITEK? 2 AST-YS07 TEST KIT |
Generic Name | VITEK? 2 AST-YS07 TEST KIT |
Product Code | NGZ |
Date Received | 2017-12-11 |
Catalog Number | 414967 |
Lot Number | 2870235203 |
Device Expiration Date | 2018-07-19 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | BIOMERIEUX, INC. |
Manufacturer Address | 595 ANGLUM ROAD HAZELWOOD MO 63042 US 63042 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2017-12-11 |