MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2017-12-12 for ORS-331 SLUSH WARMER DRAPE WITH PLATE manufactured by Microtek Dominicana, S.a..
[94824731]
A sample was received for evaluation. A small burn mark/hole was observed about 16 inches from the slush plate. No lot number was provided to perform a device history record review. Based on the sample evaluation this does not appear to be the result of a personnel, process, or material issue. Melts, though uncommon, can occur when the warmer is not draped properly, not turned off by using the power button, and/or there is insufficient fluid in the warmer basin, contrary to the operations manual, product labeling, product insert and in-service presentations. Because this appears to be related to misuse by the customer, no actions are being taken at this time.
Patient Sequence No: 1, Text Type: N, H10
[94824732]
Company rep reported that (b)(6) is stating that they found about an inch of saline in the bottom of the warmer after a surgical case. The hospital received the drape through a cardinal kit. No patient injury or treatment was reported for the incident.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 8043817-2017-00033 |
MDR Report Key | 7110751 |
Report Source | COMPANY REPRESENTATIVE,HEALTH |
Date Received | 2017-12-12 |
Date of Report | 2017-11-03 |
Date Mfgr Received | 2017-11-03 |
Date Added to Maude | 2017-12-12 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MRS JENNIFER MRKVICKA |
Manufacturer Street | 1 ECOLAB PLACE |
Manufacturer City | ST. PAUL MN 55102 |
Manufacturer Country | US |
Manufacturer Postal | 55102 |
Manufacturer Phone | 6512502237 |
Manufacturer G1 | MICROTEK DOMINICANA |
Manufacturer Street | ZONA FRANCA NO. 2 |
Manufacturer City | LA ROMANA, |
Manufacturer Country | DR |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | ORS-331 SLUSH WARMER DRAPE WITH PLATE |
Generic Name | FLUID WARMING AND SLUSH DRAPES |
Product Code | LHC |
Date Received | 2017-12-12 |
Returned To Mfg | 2017-11-09 |
Model Number | ORS-331 |
Lot Number | UNKNOWN |
Operator | HEALTH PROFESSIONAL |
Device Availability | R |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | MICROTEK DOMINICANA, S.A. |
Manufacturer Address | ZONA FRANCA NO. 2 LA ROMANA, DR |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2017-12-12 |