PHILIPS PAGEWRITER TC-70 CARDIOGRAPH MACHINES PAGE WRITER TC-70

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-12-12 for PHILIPS PAGEWRITER TC-70 CARDIOGRAPH MACHINES PAGE WRITER TC-70 manufactured by Philips.

Event Text Entries

[94850459] Time stamps recorded on the philips 12 lead ecg carts can be inaccurate by one hour. The time associated with 12 lead ecgs seems to be affected by the daylights savings and a synchronization errors with a main server. Philips continues to deny previous knowledge of this issue and will not provide documentation to warn our clinicians about the time associated with a 12 lead recording fluctuating between standard time and daylight savings time. While at the same time philips states that they have a software patch that has been in the works for over a month, designed to address the issue.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5073928
MDR Report Key7112058
Date Received2017-12-12
Date of Report2017-12-11
Date of Event2017-12-11
Date Added to Maude2017-12-13
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NamePHILIPS PAGEWRITER TC-70 CARDIOGRAPH MACHINES
Generic NameECG MACHINES
Product CodeDPS
Date Received2017-12-12
Model NumberPAGE WRITER TC-70
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrI
Device Sequence No1
Device Event Key0
ManufacturerPHILIPS


Patients

Patient NumberTreatmentOutcomeDate
10 2017-12-12

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.