EXTENDED CARE SUCTION SWAB KIT WITH BIOTENE (DENTIP ORAL SWAB) MDS096550

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-12-13 for EXTENDED CARE SUCTION SWAB KIT WITH BIOTENE (DENTIP ORAL SWAB) MDS096550 manufactured by Medline Industries.

Event Text Entries

[94697483]
Patient Sequence No: 1, Text Type: N, H10


[94697484] Oral swab stick was used to complete oral care on patient. When rn removed the swab, the sponge end remained in the back of the patient's mouth. Rn required assistance to hold patient's head still, use a light and extract the sponge end with tweezers from the back of the patient's mouth. Patient is intubated and ventilated. Swab was removed with no damage/harm to the patient. The swab from this event was discarded, but the 3 remaining dentip oral swabs were retained as well as the outer packaging for the oral pack. One oral swab was opened for investigation by clinical engineering. The adhesive between the plastic stick and sponge did not appear to have any gaps initially. With a small amount of rotating and pulling force for approx. 30 seconds, the sponge became detached from the plastic handle. Images of the handle can be viewed under attachments. It appears that adhesive is only present on one side of the plastic handle. Per site reporter: manufacturer representative was emailed, awaiting response. We have the remaining dentip swabs from the same pack, but not the broken swab from this event. Swabs and outer packaging will be given directly to medline rep.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number7112368
MDR Report Key7112368
Date Received2017-12-13
Date of Report2017-11-14
Date of Event2017-11-06
Report Date2017-11-14
Date Reported to FDA2017-11-14
Date Reported to Mfgr2017-11-14
Date Added to Maude2017-12-13
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameEXTENDED CARE SUCTION SWAB KIT WITH BIOTENE (DENTIP ORAL SWAB)
Generic NameORAL CARE, SUCTION SWAB KIT
Product CodeEHZ
Date Received2017-12-13
Catalog NumberMDS096550
Lot Number17-228
ID NumberF5 13017
Device Expiration Date2019-03-01
OperatorNURSE
Device AvailabilityY
Device Age0 YR
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMEDLINE INDUSTRIES
Manufacturer AddressTHREE LAKES DRIVE NORTHFIELD IL 60093 US 60093


Patients

Patient NumberTreatmentOutcomeDate
10 2017-12-13