TRIAGE D-DIMER PANEL 98100EU

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer,foreign,health profe report with the FDA on 2017-12-15 for TRIAGE D-DIMER PANEL 98100EU manufactured by Quidel Cardiovascular Inc..

Event Text Entries

[95733642] Customer's complaint was not replicated with in-house testing of retain lot w62880rb. No issues with analye recovery were observed and the product performed as expected. Manufacturing batch records were reviewed and found that the lot met release specifications. Based on the information available, there is no indicatoin of a product deficiency and no corrective action is required.
Patient Sequence No: 1, Text Type: N, H10


[95733643] The customer provided information regarding the following event occurring on one patient: patient a (all results on same patient same day and collection time, only difference is that pathcare does d-dimer on the green top tube heparinised blood) maxi medical lab triage d-dimer - 875ng/ml (reference range < 500ng/ml) pathcare d-dimer result - 0. 42ug/ml (reference range < 0. 5 ug/ml) customer did not know the methodology performed at pathcare site. There are no clinical details available for the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2024674-2017-00007
MDR Report Key7121925
Report SourceCONSUMER,FOREIGN,HEALTH PROFE
Date Received2017-12-15
Date of Report2017-12-15
Date of Event2017-11-10
Date Mfgr Received2017-11-17
Date Added to Maude2017-12-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJESSICA PERROTTE
Manufacturer Street9975 SUMMERS RIDGE ROAD
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8588053573
Manufacturer G1QUIDEL CARDIOVASCULAR INC.
Manufacturer Street9975 SUMMERS RIDGE ROAD
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal Code92121
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRIAGE D-DIMER PANEL
Generic NameFIBRIN SPLIT PRODUCTS, PRODUCT CODE: GHH
Product CodeGHH
Date Received2017-12-15
Model Number98100EU
Lot NumberW62880RB
Device Expiration Date2017-11-27
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerQUIDEL CARDIOVASCULAR INC.
Manufacturer Address9975 SUMMERS RIDGE ROAD SAN DIEGO CA 92121 US 92121


Patients

Patient NumberTreatmentOutcomeDate
10 2017-12-15

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.