MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2017-12-15 for OPMI PENTERO 000000-0480-801 manufactured by Carl Zeiss Meditec Ag (oberkochen).
[95824452]
Czmi made multiple attempts to obtain information regarding the surgery and the patient outcome. A response from hcp has not been received to date.
Patient Sequence No: 1, Text Type: N, H10
[95824453]
The health care professional (hcp) reported that during a procedure the lamp of the opmi pentero system shut off and prompted an error message of "error defective lamp, 1282/-1157". A czmi technical support engineer (tse) advised hcp to switch to the backup lamp; however, the problem returned in a minute or two. The tse then advised hcp to switch to a back up microscope, if possible. It is unknown if microscope was actually swapped out during the procedure, as well as the outcome of the patient.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 9615010-2017-00018 |
MDR Report Key | 7123844 |
Report Source | HEALTH PROFESSIONAL,USER FACI |
Date Received | 2017-12-15 |
Date of Report | 2017-11-16 |
Date of Event | 2017-11-16 |
Date Mfgr Received | 2017-11-16 |
Device Manufacturer Date | 2008-12-10 |
Date Added to Maude | 2017-12-15 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | BIOMEDICAL ENGINEER |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MR. WILLIAM GUSTAFSON |
Manufacturer Street | 5160 HACIENDA DRIVE |
Manufacturer City | DUBLIN CA 94568 |
Manufacturer Country | US |
Manufacturer Postal | 94568 |
Manufacturer Phone | 9255574689 |
Manufacturer G1 | CARL ZEISS MEDITEC AG (OBERKOCHEN) |
Manufacturer Street | RUDOLF-EBER-STRASSE 11 |
Manufacturer City | OBERKOCHEN, BADEN-WUERTTEMBERG 73447 |
Manufacturer Country | GM |
Manufacturer Postal Code | 73447 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | OPMI PENTERO |
Generic Name | MICROSCOPE, SURGICAL, GENERAL & PLASTIC SURGERY |
Product Code | FSO |
Date Received | 2017-12-15 |
Model Number | NA |
Catalog Number | 000000-0480-801 |
Lot Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | CARL ZEISS MEDITEC AG (OBERKOCHEN) |
Manufacturer Address | RUDOLF-EBER-STRASSE 11 OBERKOCHEN, BADEN-WUERTTEMBERG 73447 GM 73447 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Other | 2017-12-15 |