ABBOTT VANCOMYCIN ASSAY 6E44-21

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a other report with the FDA on 2017-12-18 for ABBOTT VANCOMYCIN ASSAY 6E44-21 manufactured by Microgenics Corporation.

Event Text Entries

[95772464] Customer reported a discrepant vancomycin patient result of <1. 1 on a patient who was receiving vancomycin. It was retested on other architect in the lab and the result was repeatable, sample repeated in 1:4 dilution and read 14. 4. Sample was aliquoted and sent to a reference lab and was tested by the siemens vanco assay and got a result of 32. 0 ng/ml. The problem started on (b)(6) 2017, customer had 4 random draws and 2 draws were done after medication was given between the dates (b)(6). The same patient draw from (b)(6) 2017 the patient result was 16. 4. No issues with other assays qc all in range and the retest of the proficiency sample produced repeatable results from the original testing. Customer concluded that the sample is the cause of the problem.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3010939897-2017-00003
MDR Report Key7124396
Report SourceOTHER
Date Received2017-12-18
Date of Report2017-12-14
Date of Event2017-11-14
Date Mfgr Received2017-11-21
Device Manufacturer Date2017-03-15
Date Added to Maude2017-12-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR PATRICK LYNCH
Manufacturer Street46500 KATO ROAD
Manufacturer CityFREMONT CA 94538
Manufacturer CountryUS
Manufacturer Postal94538
Manufacturer Phone5109795000
Manufacturer G1MICROGENICS CORPORATION
Manufacturer Street46500 KATO ROAD
Manufacturer CityFREMONT CA 94538
Manufacturer CountryUS
Manufacturer Postal Code94538
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameABBOTT VANCOMYCIN ASSAY
Generic NameVANCOMYCIN TEST SYSTEM
Product CodeLEH
Date Received2017-12-18
Model Number6E44-21
Catalog Number6E44-21
Lot Number39468FD01
Device Expiration Date2018-09-15
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMICROGENICS CORPORATION
Manufacturer Address46500 KATO ROAD FREMONT CA 94538 US 94538


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-12-18

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.