PROVOX LUNA ADHESIVE 8014

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-12-19 for PROVOX LUNA ADHESIVE 8014 manufactured by Atos Medical Ab.

Event Text Entries

[95811757] Investigation: no product was returned from customer. Provox luna adhesive, ref 8014, lot number was not available. Since no product was returned we cannot determine the reason why the adhesive had covered the stoma. Therefore this investigation is only theoretical. The hydrogel used in luna adhesive is known to be a material that absorbs moist/sweat very well and still having the adhesion to the skin. But if the water content is very high or very low the adhesion to the skin will decrease. In this complaint the user had the flu and much moist/sweat. This might be the reason to why the adhesive became loose and covered the stoma. The hydrogel material has a water activity of 53-61% according to the technical data sheet. If the adhesive is kept outside the intended diffusion sealed plastic bag the adhesive gets a lower adhesion because of the lower water content. According to luna ifu, contraindications: "the product shall not be used by patients with a decreased level of consciousness, patients with reduced mobility of the arms and/or hands, or patients who are unable to remove the device themselves. " -the user should always be able to remove the adhesive by himself/herself. The user has followed the instructions before the adhesive was applied. No body lotion or creams were used. No product error.
Patient Sequence No: 1, Text Type: N, H10


[95811758] This is the information that was received from the atos medical local representative: husband followed instructions and placed luna onto his wife. Patient woke in the middle of the night and the base plate had covered her stoma. After further discussions the patients husband informed me that they both had flu and were experiencing night sweats. Additional info:? Was the skin cleaned before the adhesive was applied? Yes, husband cleans the skin every time with atos wipes, he has been doing this for 14 years and does everything by the book. ? If the skin was cleaned what cleaning products was used? Atos wipes. ? Was body lotion or other moisturizing creams used on the skin around the stoma? No - absolutely definitely no lotions or creams used. Pt didn't have radiotherapy so skin is in very good condition. ? Was this the first time the customer used luna or has she used it before? Any previous experience? This was the first time the patient had tried luna, she has had no other experiences. No product will be returned for investigation
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8032044-2017-00014
MDR Report Key7127936
Date Received2017-12-19
Date of Report2017-11-28
Date of Event2017-11-28
Date Mfgr Received2017-11-28
Date Added to Maude2017-12-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. KAROLINA NILSSON
Manufacturer StreetKRAFTGATAN 8
Manufacturer CityHORBY, SE-24222
Manufacturer CountrySW
Manufacturer PostalSE-24222
Manufacturer G1ATOS MEDICAL AB
Manufacturer StreetKRAFTGATAN 8
Manufacturer CityHORBY, SE-24222
Manufacturer CountrySW
Manufacturer Postal CodeSE-24222
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NamePROVOX LUNA ADHESIVE
Generic NameSTOMAL APPLIANCE SKIN-ADHERENT PATCH, NON-STERILE
Product CodeKOX
Date Received2017-12-19
Catalog Number8014
Lot NumberUNKNOWN
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerATOS MEDICAL AB
Manufacturer AddressKRAFTGATAN 8 HORBY, SE-24222 SW SE-24222


Patients

Patient NumberTreatmentOutcomeDate
101. Deathisabilit 2017-12-19