RUBELLA IGG - IGG ANTIBODIES TO RUBELLA VIRUS 04618793190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-12-20 for RUBELLA IGG - IGG ANTIBODIES TO RUBELLA VIRUS 04618793190 manufactured by Roche Diagnostics.

Event Text Entries

[96032126] (b)(4). (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[96032127] The customer received questionable results for elecsys rubella igg immunoassay when they changed lot numbers of reagents. The lab switched from lot 206194 to 267421 and noticed that some patients which were slightly positive at 14-15 iu/ml with the old lot had negative results with the new lot. Specific patient data was requested but was not provided. The erroneous results were reported outside of the laboratory. There was no allegation of an adverse event. The customer used a cobas 8000 e 602 module. The serial number was requested but was not provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2017-03045
MDR Report Key7132965
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-12-20
Date of Report2018-01-11
Date of Event2017-11-08
Date Mfgr Received2017-12-12
Date Added to Maude2017-12-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameRUBELLA IGG - IGG ANTIBODIES TO RUBELLA VIRUS
Generic NameENZYME LINKED IMMUNOABSORBENT ASSAY, RUBELLA
Product CodeLFX
Date Received2017-12-20
Model NumberNA
Catalog Number04618793190
Lot Number267421
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2017-12-20