[96032290] Unique identifier (udi)#: (b)(4). This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[96032291] The customer experienced an issue where the elecsys free psa immunoassay results were higher than the elecsys total psa immunoassay results for two patients. The customer used the cobas 8000 e 602 module serial number (b)(4). For patient 1 the free psa result was 2. 19 ng/ml and the total psa result was 0. 07 ng/ml. For patient 2 on (b)(6) 2017 the free psa result was 3. 7ng/ml and the total psa result was 0. 065 ng/ml. On (b)(6) 2017 the patient 2 sample was repeated with an automatic dilution and the free psa result was 3. 53 ng/ml. The patient 2 sample was tested using the siemens method and the free psa result was 0. 02 ng/ml and the total psa result was 3. 24 ng/ml. The customer believed the results from the siemens method. The results were reported outside of the laboratory. There was no allegation of an adverse event. This medwatch will cover the erroneous free psa results on the e 602 module. Patient identifier (b)(6) for information on the erroneous total psa results from the e 602 module.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2017-03051
MDR Report Key7133019
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-12-20
Date of Report2018-01-18
Date of Event2017-11-17
Date Mfgr Received2017-12-11
Date Added to Maude2017-12-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameFREE PSA
Generic NameTEST,PROSTATE SPECIFIC ANTIGEN,FREE,(NONCOMPLEXED) TO DISTINGUISH PROSTATE CA
Product CodeMTG
Date Received2017-12-20
Model NumberNA
Catalog Number03289788190
Lot Number20987700
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2017-12-20