ELECSYS FREE PSA IMMUNOASSAY 03289788160

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2017-12-20 for ELECSYS FREE PSA IMMUNOASSAY 03289788160 manufactured by Roche Diagnostics.

Event Text Entries

[95903516] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[95903517] The customer complained of questionable results for 1 patient sample tested for elecsys free psa immunoassay elecsys (psa free) total psa immunoassay (psa total) on a cobas 8000 e 602 module. (b)(4). The initial psa free result was 0. 3 ng/ml which was greater than the initial psa total result of 0. 06 ng/ml. There were no flags present on the initial results. The customer repeated the sample for both psa free and psa total on both the same analyzer and on another analyzer, running the sample both undiluted and diluted, and all the results obtained matched (no specific values provided). The erroneous results were reported outside of the laboratory. There was no adverse event. The customer stated that all qc results have been within range. The field engineering specialist was not able to find a cause. He checked the instrument and found no issues. He ran performance and precision testing which had acceptable results. Psa total qc testing was performed which passed. (b)(6).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2017-03058
MDR Report Key7133310
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2017-12-20
Date of Report2018-02-12
Date of Event2017-12-09
Date Mfgr Received2017-12-11
Date Added to Maude2017-12-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameELECSYS FREE PSA IMMUNOASSAY
Generic NameTEST,PROSTATE SPECIFIC ANTIGEN,FREE,(NONCOMPLEXED) TO DISTINGUISH PROSTATE CA
Product CodeMTG
Date Received2017-12-20
Model NumberNA
Catalog Number03289788160
Lot Number24314201
ID NumberNA
Device Expiration Date2018-09-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457