TRUCLEAR 5 OP HYSTERSCOPE 72202976

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-12-20 for TRUCLEAR 5 OP HYSTERSCOPE 72202976 manufactured by Covidien Lp Llc North Haven.

Event Text Entries

[95664533] Post market vigilance (pmv) led an evaluation of one device. A visual inspection was performed and showed distal tip damage, dent in the sheath, and a damaged leaky stopcock. Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all medtronic quality release specifications at the time of manufacture. Analysis concluded there were no assembly component related failures. The instructions for use (ifu) recommends?? Careful inspection of the operative hysteroscope before and after the procedure for possible signs of damage. Immediate detection and repair of minor damage will extend the life of the operative hysteroscope.? Damage may occur if the hysteroscope is handled roughly during use. Excessive forces may result in traumatic insertion, chipping particles of the optical glass, or other damage to the unit. Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.
Patient Sequence No: 1, Text Type: N, H10


[95664534] According to the reporter, the used device had foggy lens. No additional information given.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1219930-2017-10252
MDR Report Key7135815
Date Received2017-12-20
Date of Report2017-12-20
Date of Event2017-05-25
Date Mfgr Received2017-12-07
Date Added to Maude2017-12-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSHARON MURPHY
Manufacturer Street60 MIDDLETOWN AVE
Manufacturer CityNORTH HAVEN CT 06473
Manufacturer CountryUS
Manufacturer Postal06473
Manufacturer Phone2034925267
Manufacturer G1COVIDIEN LP LLC NORTH HAVEN
Manufacturer Street195 MCDERMOTT RD
Manufacturer CityNORTH HAVEN 06473
Manufacturer CountryUS
Manufacturer Postal Code06473
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameTRUCLEAR 5 OP HYSTERSCOPE
Generic NameHYSTEROSCOPE ACCESSORIES
Product CodeNWW
Date Received2017-12-20
Returned To Mfg2017-06-06
Model Number72202976
Catalog Number72202976
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCOVIDIEN LP LLC NORTH HAVEN
Manufacturer Address195 MCDERMOTT RD NORTH HAVEN 06473 US 06473


Patients

Patient NumberTreatmentOutcomeDate
10 2017-12-20

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.